期刊论文详细信息
Journal of Cardiovascular Magnetic Resonance
Myocardial arterial spin labeling perfusion imaging with improved sensitivity
Technical Notes
Terrence R Jao1  Hung Phi Do2  Krishna S Nayak3 
[1] Department of Biomedical Engineering, University of Southern California, Los Angeles, CA, USA;Department of Physics and Astronomy, University of Southern California, 3740 McClintock Ave, EEB 400, 90089-2564, Los Angeles, CA, USA;Ming Hsieh Department of Electrical Engineering, University of Southern California, Los Angeles, CA, USA;
关键词: Arterial spin labeling;    Myocardial perfusion;    Physiological noise;    Sensitivity;    Parallel imaging;   
DOI  :  10.1186/1532-429X-16-15
 received in 2013-09-18, accepted in 2014-01-22,  发布年份 2014
来源: Springer
PDF
【 摘 要 】

BackgroundMyocardial arterial spin labeling (ASL) is a noninvasive MRI based technique that is capable of measuring myocardial blood flow (MBF) in humans. It suffers from poor sensitivity to MBF due to high physiological noise (PN). This study aims to determine if the sensitivity of myocardial ASL to MBF can be improved by reducing image acquisition time, via parallel imaging.MethodsMyocardial ASL scans were performed in 7 healthy subjects at rest using flow-sensitive alternating inversion recovery (FAIR) tagging and balanced steady state free precession (SSFP) imaging. Sensitivity encoding (SENSE) with a reduction factor of 2 was used to shorten each image acquisition from roughly 300 ms per heartbeat to roughly 150 ms per heartbeat. A paired Student’s t-test was performed to compare measurements of myocardial blood flow (MBF) and physiological noise (PN) from the reference and accelerated methods.ResultsThe measured PN (mean ± standard deviation) was 0.20 ± 0.08 ml/g/min for the reference method and 0.08 ± 0.05 ml/g/min for the accelerated method, corresponding to a 60% reduction. PN measured from the accelerated method was found to be significantly lower than that of the reference method (p = 0.0059). There was no significant difference between MBF measured from the accelerated and reference ASL methods (p = 0.7297).ConclusionsIn this study, significant PN reduction was achieved by shortening the acquisition window using parallel imaging with no significant impact on the measured MBF. This indicates an improvement in sensitivity to MBF and may also enable the imaging of subjects with higher heart rates and imaging during systole.

【 授权许可】

CC BY   
© Do et al.; licensee BioMed Central Ltd. 2014

【 预 览 】
附件列表
Files Size Format View
RO202311104779423ZK.pdf 525KB PDF download
【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  文献评价指标  
  下载次数:0次 浏览次数:3次