会议论文详细信息
12th Congress of Indonesian Soc. for Biochemistry and Molecular Biology in Conjunction With The 2nd Int. Conf. "Collaboration Seminar of Chemistry and Industry
Carbon Paste Electrode Modified Imprinted Zeolite as a Selective Sensor for Creatine Analysis by Potentiometry
生物科学;化学
Athiroh, A.^1 ; Fadillah, T.^1 ; Damayanti, D.F.^1 ; Widati, A.A.^1 ; Abdulloh, A.^1 ; Khasanah, M.^1
Chemistry Department, Faculty of Science and Technology, Universitas Airlangga Kampus, Surabaya
60115, Indonesia^1
关键词: Carbon paste electrode;    Coefficient of variation;    Creatine;    Lower detection limit;    Modified electrodes;    Optimum performance;    Potentiometry;    Selective sensors;   
Others  :  https://iopscience.iop.org/article/10.1088/1755-1315/217/1/012003/pdf
DOI  :  10.1088/1755-1315/217/1/012003
学科分类:生物科学(综合)
来源: IOP
PDF
【 摘 要 】

Carbon paste electrode modified imprinted zeolite have been developed as a sensor to analyze creatine by potentiometry. Zeolite used in this study was TS-1 type zeolite that was synthesized with mole ratio of TEOS, TBOT, TPAOH, and H2O of 1:0.017:0.24:21.2. Imprinted zeolite (IZ) was synthesized by mole ratio of creatine/Si of 0.0306. The developed electrode showed the optimum performance on pH 5, range of measurement of 10-9-10-4 M, linearity of 0.973, and Nernst factor of 27.31 mV/decade. The upper and lower detection limits were 5.5×10-5 M and 1.3×10-8 M, respectively. The electrode showed a good precision which is expressed by coefficient of variation of 1.13%-1.43%. The response time of the electrode in 10-9-10-4 M creatine was 82-199 seconds, while its life time was 7 weeks (83 times used). Urea, glucose, and uric acid in various concentrations did not interfere on creatine analysis by potentiometry using the modified electrode. Application of the modified electrode in two serum samples with spiking technique resulted in recovery of 76.8% and 81.8%. Comparative testing with the standard clinical laboratory method showed accuracy of 73%.

【 预 览 】
附件列表
Files Size Format View
Carbon Paste Electrode Modified Imprinted Zeolite as a Selective Sensor for Creatine Analysis by Potentiometry 216KB PDF download
  文献评价指标  
  下载次数:42次 浏览次数:35次