期刊论文详细信息
Plasma
Analysis and Mitigation of Pulse-Pile-Up Artifacts in Plasma Pulse-Height X-ray Spectra
article
Taosif. Ahsan1  Charles P. S. Swanson2  Chris Galea3  Sangeeta P. Vinoth2  Tony Qian2  Tal Rubin2  Samuel A. Cohen2 
[1] Physics Department, Princeton University, Princeton;Princeton Plasma Physics Laboratory, Princeton University, Princeton;Princeton Fusion Systems
关键词: X-ray;    pulse pile up;    energy analyzers;    plasma;    2-photon;    dead time;    count rate;   
DOI  :  10.3390/plasma6010006
学科分类:泌尿医学
来源: mdpi
PDF
【 摘 要 】

Pulse pile-up in pulse-height energy analyzers increases when the incident rate of pulses increases relative to the inverse of the dead time per pulse of the detection system. Changes in the observed energy distributions with incident rate and detector-electronics-formed pulse shape then occur. We focus on weak high energy tails in X-ray spectra, important for measurements on partially ionized, warm (50–500 eV average electron energy), pure hydrogen plasma. A first-principles two-photon pulse-pile-up model is derived specific to trapezoidal-shaped pulses; quantitative agreement is found between the measurements and the model’s predictions. The model is then used to diagnose pulse-pile-up tail artifacts and mitigate them in relatively low count-rate spectra.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202307010002699ZK.pdf 3890KB PDF download
  文献评价指标  