Sensors & Transducers | |
Cholesterol Biosensor Based on Polyvinyl Formal Membrane Bound Cholesterol Esterase and Oxidase | |
Ashok Kumar1  Suman2  | |
[1] Institute of Genomics and Integrative Biology, Mall Road, Delhi-110007, India;Shriram Institute for Industrial Research, 19-University Road, Delhi-110007, India; | |
关键词: Cholesterol esterase; Cholesterol oxidase; Cholesterol biosensor; PVA membrane; | |
DOI : | |
来源: DOAJ |
【 摘 要 】
Cholesterol biosensor was fabricated by co-immobilizing cholesterol esterase and cholesterol oxidase in polyvinyl formal membrane and by mounting over the sensing part of oxygen electrode. The immobilized enzymes membrane was characterized by atomic force microscopy. The biosensor showed the linearity for 100-450 mg dl-1 cholesterol acetate. The minimum detection limit of the cholesterol was 50 mg dl-1. The effects of temperature, pH and the stability of immobilized enzymes were also studied. A good correlation was found among cholesterol values obtained by colorimetric, Enzo-kit and auto analyzer method. The shelf-life of the biosensor was more than three month at 4ºC and enzyme membrane can be re-used 20 times without any significant loss in enzyme activity.
【 授权许可】
Unknown