Motion compensated whole-heart coronary cardiovascular magnetic resonance angiography using focused navigation (fNAV)
Giulia Rossi1  Christopher W. Roy1  John Heerfordt2  Davide Piccini2  Matthias Stuber3  Anna Giulia Pavon4  Juerg Schwitter5 
[1] Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Rue de Bugnon 46, BH-7-84, 1011, Lausanne, Switzerland;Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Rue de Bugnon 46, BH-7-84, 1011, Lausanne, Switzerland;Advanced Clinical Imaging Technology (ACIT), Siemens Healthcare AG, Lausanne, Switzerland;Department of Radiology, Lausanne University Hospital (CHUV) and University of Lausanne (UNIL), Rue de Bugnon 46, BH-7-84, 1011, Lausanne, Switzerland;Center for Biomedical Imaging (CIBM), Lausanne, Switzerland;Division of Cardiology, Lausanne University Hospital (CHUV), Lausanne, Switzerland;Division of Cardiology, Lausanne University Hospital (CHUV), Lausanne, Switzerland;Director CMR-Center, Lausanne University Hospital (CHUV), Lausanne, Switzerland;Faculty of Biology and Medicine, University of Lausanne (UNIL), Lausanne, Switzerland;
关键词: Coronary magnetic resonance angiography;    Motion correction;    Whole heart;    Free-breathing;   
DOI  :  10.1186/s12968-021-00717-4
来源: Springer
PDF
【 摘 要 】

BackgroundRadial self-navigated (RSN) whole-heart coronary cardiovascular magnetic resonance angiography (CCMRA) is a free-breathing technique that estimates and corrects for respiratory motion. However, RSN has been limited to a 1D rigid correction which is often insufficient for patients with complex respiratory patterns. The goal of this work is therefore to improve the robustness and quality of 3D radial CCMRA by incorporating both 3D motion information and nonrigid intra-acquisition correction of the data into a framework called focused navigation (fNAV).MethodsWe applied fNAV to 500 data sets from a numerical simulation, 22 healthy subjects, and 549 cardiac patients. In each of these cohorts we compared fNAV to RSN and respiratory resolved extradimensional golden-angle radial sparse parallel (XD-GRASP) reconstructions of the same data. Reconstruction times for each method were recorded. Motion estimate accuracy was measured as the correlation between fNAV and ground truth for simulations, and fNAV and image registration for in vivo data. Percent vessel sharpness was measured in all simulated data sets and healthy subjects, and a subset of patients. Finally, subjective image quality analysis was performed by a blinded expert reviewer who chose the best image for each in vivo data set and scored on a Likert scale 0–4 in a subset of patients by two reviewers in consensus.ResultsThe reconstruction time for fNAV images was significantly higher than RSN (6.1 ± 2.1 min vs 1.4 ± 0.3, min, p < 0.025) but significantly lower than XD-GRASP (25.6 ± 7.1, min, p < 0.025). Overall, there is high correlation between the fNAV and reference displacement estimates across all data sets (0.73 ± 0.29). For simulated data, healthy subjects, and patients, fNAV lead to significantly sharper coronary arteries than all other reconstruction methods (p < 0.01). Finally, in a blinded evaluation by an expert reviewer fNAV was chosen as the best image in 444 out of 571 data sets (78%; p < 0.001) and consensus grades of fNAV images (2.6 ± 0.6) were significantly higher (p < 0.05) than uncorrected (1.7 ± 0.7), RSN (1.9 ± 0.6), and XD-GRASP (1.8 ± 0.8).ConclusionfNAV is a promising technique for improving the quality of RSN free-breathing 3D whole-heart CCMRA. This novel approach to respiratory self-navigation can derive 3D nonrigid motion estimations from an acquired 1D signal yielding statistically significant improvement in image sharpness relative to 1D translational correction as well as XD-GRASP reconstructions. Further study of the diagnostic impact of this technique is therefore warranted to evaluate its full clinical utility.

【 授权许可】

CC BY   

【 预 览 】
附件列表
Files Size Format View
RO202107029031325ZK.pdf 4921KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:6次