Journal of Cardiovascular Magnetic Resonance | |
Free-breathing myocardial T2* mapping using GRE-EPI and automatic Non-rigid motion correction | |
Research | |
Nathan Lamba1  Juliana Serafim da Silveira1  Subha V. Raman2  Orlando P. Simonetti3  Sharath Subramanian4  Marie-Pierre Jolly5  George Mathew6  David N. Firmin7  Dudley J. Pennell7  Ning Jin8  | |
[1] Davis Heart & Lung Research Institute, The Ohio State University, Columbus, OH, USA;Davis Heart & Lung Research Institute, The Ohio State University, Columbus, OH, USA;Department of Internal Medicine – Division of Cardiovascular Medicine, The Ohio State University, Columbus, OH, USA;Davis Heart & Lung Research Institute, The Ohio State University, Columbus, OH, USA;Department of Internal Medicine – Division of Cardiovascular Medicine, The Ohio State University, Columbus, OH, USA;Department of Biomedical Engineering, The Ohio State University, Columbus, OH, USA;Department of Radiology, The Ohio State University, Columbus, OH, USA;Department of Internal Medicine – Division of Cardiovascular Medicine, The Ohio State University, Columbus, OH, USA;Imaging & Computer Vision, Siemens Corporation, Corporate Technology, Princeton, NJ, USA;NIHR Cardiovascular Biomedical Research Unit, Royal Brompton and Harefield NHS Trust, London, UK;NIHR Cardiovascular Biomedical Research Unit, Royal Brompton and Harefield NHS Trust, London, UK;National Heart and Lung Institute, Imperial College, London, UK;Siemens Medical Solutions USA, Inc, 460 West 12th Ave, Room 311, OH 43210, Columbus, OH, USA; | |
关键词: Cardiac; T2*; Mapping; Free-breathing; Iron overload; GRE-EPI; Motion correction; | |
DOI : 10.1186/s12968-015-0216-z | |
received in 2015-08-05, accepted in 2015-12-09, 发布年份 2015 | |
来源: Springer | |
【 摘 要 】
BackgroundMeasurement of myocardial T2* is becoming widely used in the assessment of patients at risk for cardiac iron overload. The conventional breath-hold, ECG-triggered, segmented, multi-echo gradient echo (MGRE) sequence used for myocardial T2* quantification is very sensitive to respiratory motion and may not be feasible in patients who are unable to breath-hold. We propose a free-breathing myocardial T2* mapping approach that combines a single-shot gradient-echo echo-planar imaging (GRE-EPI) sequence for T2*-weighted image acquisition with automatic non-rigid motion correction (MOCO) of respiratory motion between single-shot images.MethodsECG-triggered T2*-weighted images at different echo times were acquired by a black-blood, single-shot GRE-EPI sequence during free-breathing. A single image at a single TE is acquired in each heartbeat. Automatic non-rigid MOCO was applied to correct for in-plane respiratory motion before pixel-wise T2* mapping. In a total of 117 patients referred for clinical cardiac magnetic resonance exams, the free-breathing MOCO GRE-EPI sequence was compared to the breath-hold segmented MGRE approach. Image quality was scored independently by 2 experienced observers blinded to the particular image acquisition strategy. T2* measurements in the interventricular septum and in the liver were compared for the two methods in all cases with adequate image quality.ResultsT2* maps were acquired in all 117 patients using the breath-hold MGRE and the free-breathing MOCO GRE-EPI approaches, including 8 patients with myocardial iron overload and 25 patients with hepatic iron overload. The mean image quality of the free-breathing MOCO GRE-EPI images was scored significantly higher than that of the breath-hold MGRE images by both reviewers. Out of the 117 studies, 21 breath-hold MGRE studies (17.9 % of all the patients) were scored to be less than adequate or very poor by both reviewers, while only 2 free-breathing MOCO GRE-EPI studies were scored to be less than adequate image quality. In a comparative evaluation of the images with at least adequate quality, the intra-class correlation coefficients for myocardial and liver T2* were 0.868 and 0.986 respectively (p < 0.001), indicating that the T2* measured by breath-hold MGRE and free-breathing MOCO GRE-EPI were in close agreement. The coefficient of variation between the breath-hold and free-breathing approaches for myocardial and liver T2* were 9.88 % and 9.38 % respectively. Bland-Altman plots demonstrated good absolute agreement of T2* in the interventricular septum and the liver from the free-breathing and breath-hold approaches (mean differences -0.03 and 0.16 ms, respectively).ConclusionThe free-breathing approach described for T2* mapping using MOCO GRE-EPI enables accurate myocardial and liver T2* measurements and is insensitive to respiratory motion.
【 授权许可】
CC BY
© Jin et al. 2015
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202311105491074ZK.pdf | 2344KB | download |
【 参考文献 】
- [1]
- [2]
- [3]
- [4]
- [5]
- [6]
- [7]
- [8]
- [9]
- [10]
- [11]
- [12]
- [13]
- [14]
- [15]
- [16]
- [17]
- [18]
- [19]
- [20]
- [21]
- [22]
- [23]
- [24]
- [25]
- [26]
- [27]
- [28]
- [29]
- [30]
- [31]
- [32]
- [33]
- [34]
- [35]
- [36]
- [37]
- [38]
- [39]
- [40]
- [41]