会议论文详细信息
8th International Conference on 3D Radiation Dosimetry
Development of CCD-based optical computed tomography and comparison with single-beam optical CT scanner
Chang, Y.J.^1,2
Institute of Biomedical Engineering and Materials Science, Central Taiwan University of Science and Technology, Taichung, Taiwan^1
Department of Management Information Systems, Central Taiwan University of Science and Technology, Taichung, Taiwan^2
关键词: Optical computed tomography;    Optical CT;    Pass rate;    Polymer gel dosimeter;    Scanning time;    Single beam;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012060/pdf
DOI  :  10.1088/1742-6596/573/1/012060
来源: IOP
PDF
【 摘 要 】

This study reports on the development of CCD-based optical computed tomography (CT) CT-s2. A commercially available 10× fast optical computed tomography scanner (OCTOPUSTM-10X, MGS Research, Inc., Madison, CT, USA) was used for comparison. NIPAM polymer gel dosimeter was used to validate the performance of CT-s2. The gamma pass rate can reach 96.00% when using a 3% dose difference and 3 mm dose-to-agreement criteria. The results of CT-s2 are as good as those of the single-beam optical-CT scanner, but the scanning time of CT-s2 is only one-tenth of that of the single-beam optical-CT scanner.

【 预 览 】
附件列表
Files Size Format View
Development of CCD-based optical computed tomography and comparison with single-beam optical CT scanner 785KB PDF download
  文献评价指标  
  下载次数:18次 浏览次数:7次