会议论文详细信息
| 8th International Conference on 3D Radiation Dosimetry | |
| Development of CCD-based optical computed tomography and comparison with single-beam optical CT scanner | |
| Chang, Y.J.^1,2 | |
| Institute of Biomedical Engineering and Materials Science, Central Taiwan University of Science and Technology, Taichung, Taiwan^1 | |
| Department of Management Information Systems, Central Taiwan University of Science and Technology, Taichung, Taiwan^2 | |
| 关键词: Optical computed tomography; Optical CT; Pass rate; Polymer gel dosimeter; Scanning time; Single beam; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012060/pdf DOI : 10.1088/1742-6596/573/1/012060 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
This study reports on the development of CCD-based optical computed tomography (CT) CT-s2. A commercially available 10× fast optical computed tomography scanner (OCTOPUSTM-10X, MGS Research, Inc., Madison, CT, USA) was used for comparison. NIPAM polymer gel dosimeter was used to validate the performance of CT-s2. The gamma pass rate can reach 96.00% when using a 3% dose difference and 3 mm dose-to-agreement criteria. The results of CT-s2 are as good as those of the single-beam optical-CT scanner, but the scanning time of CT-s2 is only one-tenth of that of the single-beam optical-CT scanner.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Development of CCD-based optical computed tomography and comparison with single-beam optical CT scanner | 785KB |
PDF