会议论文详细信息
8th International Conference on 3D Radiation Dosimetry | |
High resolution 3D dosimetry for microbeam radiation therapy using optical CT | |
McErlean, C.^1 ; Bräuer-Krisch, E.^2 ; Adamovics, J.^3,6 ; Leach, M.O.^1,4 ; Doran, S.J.^1,5 | |
CRUK Cancer Imaging Centre, Institute of Cancer Research, London, United Kingdom^1 | |
European Synchrotron Radiation Facility, Grenoble, France^2 | |
Department of Chemistry and Biology, Rider University, Lawrenceville | |
NJ, United States^3 | |
Clinical Magnetic Resonance Research Group, Royal Marsden NHS Trust, United Kingdom^4 | |
Department of Physics, University of Surrey, Guildford, United Kingdom^5 | |
Heuris Pharma, LLC, 412 Sunset Rd., Skillman | |
NJ | |
08558, United States^6 | |
关键词: High resolution; Irradiation geometries; Microbeam Radiation Therapy; Optical computed tomography; Optical CT; Peak-to-valley dose ratios; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012032/pdf DOI : 10.1088/1742-6596/573/1/012032 |
|
来源: IOP | |
【 摘 要 】
Optical Computed Tomography (CT) is a promising technique for dosimetry of Microbeam Radiation Therapy (MRT), providing high resolution 3D dose maps. Here different MRT irradiation geometries are visualised showing the potential of Optical CT as a tool for future MRT trials. The Peak-to-Valley dose ratio (PVDR) is calculated to be 7 at a depth of 3mm in the radiochromic dosimeter PRESAGE®. This is significantly lower than predicted values and possible reasons for this are discussed.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
High resolution 3D dosimetry for microbeam radiation therapy using optical CT | 907KB | download |