Real-world examples of sensitivity failures of the 3%/3mm pass rate metric and published action levels when used in IMRT/VMAT system commissioning | |
Nelms, B.^1 ; Jarry, G.^2 ; Chan, M.^3 ; Hampton, C.^4 ; Watanabe, Y.^5 ; Feygelman, V.^6 | |
Canis Lupus LLC, Merrimac, WI 53561, United States^1 | |
Hopital Maissoneuve-Rosemont, Montreal, QC H1T 2M4, Canada^2 | |
Memorial Sloan-Kettering Cancer Center, Basking Ridge, NJ 07920, United States^3 | |
Wake Forest Baptist Medical Center, Winston-Salem, NC 27157, United States^4 | |
University of Minnesota, Minneapolis, MN 55455, United States^5 | |
Department of Radiation Oncology, Moffitt Cancer Center, Tampa, FL 33612, United States^6 | |
关键词: Acceptance testing; Delivery chain; Dose analysis; False negatives; Gamma analysis; Pass rate; Real-world; Tight tolerances; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/444/1/012086/pdf DOI : 10.1088/1742-6596/444/1/012086 |
|
来源: IOP | |
【 摘 要 】
In IMRT/VMAT system commissioning (as with any system), quality is improved by striving for tight tolerances of stringent metrics of accuracy. For 5 cases, passing rates for 3%/3mm gamma analysis were generated following the TG119 instructions. Subsequently, more stringent/sensitive criteria combined with advanced volumetric dose analysis were applied, and in each case significant systematic errors were clearly identified despite the high 3%/3mm passing rates. In 4 of 5 cases, the error was easily remedied. These real-world examples of observed "false negatives" (insensitivities) point towards the inappropriateness of the 3%/3mm gamma passing rate metric as the basis for acceptance testing/commissioning of the IMRT/VMAT delivery chain.