Quantitative Measurement of Highly Focused Ultrasound Pressure Field by Optical Shadowgraph | |
Miyasaka, R.^1 ; Harigane, S.^1 ; Yoshizawa, S.^1 ; Umemura, S.^2 | |
Department of Electrical and Communication Engineering, Tohoku University, Sendai 980-8579, Japan^1 | |
Department of Biomedical Engineering, Tohoku University, Sendai 980-8579, Japan^2 | |
关键词: High intensity focused ultrasound; Highly focused ultrasounds; Holographic diffusers; Medical ultrasound; Optical shadowgraphs; Propagation lengths; Quantitative measurement; Ultrasound pressure; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/520/1/012026/pdf DOI : 10.1088/1742-6596/520/1/012026 |
|
来源: IOP | |
【 摘 要 】
In the development of medical ultrasound techniques, fast and accurate pressure field measurement is important. The most common method to measure an ultrasound pressure field is mechanically scanning a hydrophone, which takes a long time and might disturb the acoustic field. In this study, we used an optical shadowgraph method. To perform this method quantitatively, it is important to define the optical propagation length precisely. For this purpose, a holographic diffuser was used as the imaging screen. Combined with a computed tomography (CT) algorithm, a pressure field was reconstructed, and the result was compared with that of hydrophone measurement. By using two shadowgraph data from short and long propagation lengths, the pressure field was successfully reconstructed even at a pressure level for high intensity focused ultrasound (HIFU) treatment.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Quantitative Measurement of Highly Focused Ultrasound Pressure Field by Optical Shadowgraph | 994KB | download |