New Electrostatically-Focused, UV HPD for Liquid Xenon: A Direct Comparison with APD, PMT, SiPM in an Integrated Database (08/15/05-08/14/07). | |
Aprile, E. | |
Technical Information Center Oak Ridge Tennessee | |
关键词: Xenon; Photodiodes; Comparisons; Graphs (Charts); Experimental setup; | |
RP-ID : DE2010963054 | |
学科分类:工程和技术(综合) | |
美国|英语 | |
来源: National Technical Reports Library | |
【 摘 要 】
The operation of different avalanche photodiodes (APDs) for detection of liquid xenon scintillation was investigated. Quantum efficiencies of 1.08 plus or minus 0.05 and 0.80 plus or minus 0.14 where measured for two APD's provided by Advanced Photonix Inc. (API) and Hamamatsu Photonics, respectively. Both APD types were found to reach gains above 200, although gains as low as a few tens may prove sufficient to achieve optimum performance. Compared to the API photodiodes, the Hamamatsu APDs used in these tests exhibited poor performance.