科技报告详细信息
Coordinate Standard Measurement Development.
Hanshaw, R. A.
Technical Information Center Oak Ridge Tennessee
关键词: Calibration standards;    Interferometers;    Modifications;    Reliability;    Resolution;   
RP-ID  :  DE2001751342
学科分类:工程和技术(综合)
美国|英语
来源: National Technical Reports Library
PDF
【 摘 要 】

A Shelton Precision Interferometer Base, which is used for calibration of coordinate standards, was improved through hardware replacement, software geometry error correction, and reduction of vibration effects. Substantial increases in resolution and reliability, as well as reduction in sampling time, were achieved through hardware replacement; vibration effects were reduced substantially through modification of the machine component dampening and software routines; and the majority of the machine's geometry error was corrected through software geometry error correction. Because of these modifications, the uncertainty of coordinate standards calibrated on this device has been reduced dramatically.