科技报告详细信息
Results from the Cooler and Lead Tests
Green, Michael A
Lawrence Berkeley National Laboratory
关键词: Copper;    Spectrometers;    Testing;    42;    Voltage Drop;   
DOI  :  10.2172/988858
RP-ID  :  LBNL-3926E
RP-ID  :  DE-AC02-05CH11231
RP-ID  :  988858
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

The report presents the results of testing MICE spectrometer magnet current leads on a test apparatus that combines both the copper leads and the high temperature superconducting (HTS) leads with a single Cryomech PT415 cooler and liquid helium tank. The current is carried through the copper leads from 300 K to the top of the HTS leads. The current is then carried through the HTS leads to a feed-through from the vacuum space to the inside of a liquid helium tank. The experiment allows one to measure the performance of both cooler stages along with the performance of the leads. While the leads were powered we measured the voltage drops through the copper leads, through the HTS leads, through spliced to the feed-through, through the feed-through and through the low-temperature superconducting loop that connects one lead to the other. Measurements were made using the leads that were used in spectrometer magnet 1A and spectrometer magnet 2A. These are the same leads that were used for Superbend and Venus magnets at LBNL. The IL/A for these leads was 5.2 x 10{sup 6} m{sup -1}. The leads turned out to be too long. The same measurements were made using the leads that were installed in magnet 2B. The magnet 2B leads had an IL/A of 3.3 x 10{sup 6} A m{sup -1}. This report discusses the cooler performance and the measured electrical performance of the lead circuit that contains the copper leads and the superconducting leads. All of the HTS leads that were installed in magnet 2B were current tested using this apparatus.

【 预 览 】
附件列表
Files Size Format View
988858.pdf 13254KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:16次