期刊论文详细信息
| NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS | 卷:376 |
| Upgrade of the TAMU MDM-focal plane detector with MicroMegas technology | |
| Article; Proceedings Paper | |
| Spiridon, A.1  Pollacco, E.2  Roeder, B. T.1  Saastamoinen, A.1  Chyzh, R.1  Dag, M.1  Tribble, R. E.1  Trache, L.3  Pascovici, G.3  De Oliveira, R.4  | |
| [1] Texas A&M Univ, Inst Cyclotron, College Stn, TX 77843 USA | |
| [2] CEA Saclay, IRFU, F-91191 Gif Sur Yvette, France | |
| [3] Horia Hulubei Natl Inst Phys & Nucl Engn, RO-077125 Bucharest, Romania | |
| [4] CERN, Geneva, Switzerland | |
| 关键词: Radioactive ion beams; Micro pattern gas amplifier detector; MPGAD; Spectrometers; | |
| DOI : 10.1016/j.nimb.2016.02.011 | |
| 来源: Elsevier | |
PDF
|
|
【 摘 要 】
A gridded ionization chamber used as a focal plane detector at the back of the TAMU-MDM spectrometer was modified to use MicroMegas technology for the purpose of improving energy resolution and particle identification. The upgraded system was tested in experimental conditions with several heavy-ion beams at 12 MeV/u and found to achieve resolutions between 3.2% and 4.8%. This is a significant improvement over the previous performance of 10-15% obtained using the existing, conventional ionization chambers. (C) 2016 Elsevier B.V. All rights reserved.
【 授权许可】
Free
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 10_1016_j_nimb_2016_02_011.pdf | 2165KB |
PDF