【 摘 要 】

The Phase II period performance was May 30, 2002 through May 29, 2004. This development effort was successfully completed within the period and budget allotted. The proposed design was successfully fabricated from B{sub 4}C-coated aluminum and copper film, slit and wound to form 4 mm diameter straws, cut to 100 cm in length, and threaded with resistive anode wires (20 {micro}m in diameter). This paper reports testing done with two 50-straw detector modules at the reactor of the Nuclear Science Center at Texas A&M University (TAMU NSC).

【 预 览 】
附件列表
Files Size Format View
828684.pdf 1247KB PDF download