科技报告详细信息
Benchmarks for GADRAS performance validation.
Mattingly, John K. ; Mitchell, Dean James ; Rhykerd, Charles L., Jr.
Sandia National Laboratories
关键词: Calibration;    Dimensions;    Response Functions;    Depleted Uranium;    11 Nuclear Fuel Cycle And Fuel Materials;   
DOI  :  10.2172/1031909
RP-ID  :  SAND2009-6550
RP-ID  :  AC04-94AL85000
RP-ID  :  1031909
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】

The performance of the Gamma Detector Response and Analysis Software (GADRAS) was validated by comparing GADRAS model results to experimental measurements for a series of benchmark sources. Sources for the benchmark include a plutonium metal sphere, bare and shielded in polyethylene, plutonium oxide in cans, a highly enriched uranium sphere, bare and shielded in polyethylene, a depleted uranium shell and spheres, and a natural uranium sphere. The benchmark experimental data were previously acquired and consist of careful collection of background and calibration source spectra along with the source spectra. The calibration data were fit with GADRAS to determine response functions for the detector in each experiment. A one-dimensional model (pie chart) was constructed for each source based on the dimensions of the benchmark source. The GADRAS code made a forward calculation from each model to predict the radiation spectrum for the detector used in the benchmark experiment. The comparisons between the GADRAS calculation and the experimental measurements are excellent, validating that GADRAS can correctly predict the radiation spectra for these well-defined benchmark sources.

【 预 览 】
附件列表
Files Size Format View
1031909.pdf 1630KB PDF download
  文献评价指标  
  下载次数:11次 浏览次数:56次