科技报告详细信息
Shortwave Spectroradiometer (SWS) Handbook
Pilewskie, P ; Pommier, J
Pacific Northwest National Laboratory (U.S.)
关键词: Spectrometers;    Manuals;    Wavelengths;    Collimators;    Personnel;   
DOI  :  10.2172/1020259
RP-ID  :  DOE/SC-ARM/TR-062
RP-ID  :  DE-AC05-7601830
RP-ID  :  1020259
美国|英语
来源: UNT Digital Library
PDF
【 摘 要 】
The Shortwave Spectroradiometer (SWS), measures the absolute visible and near infrared spectralradiance (units of watts per meter square per nanometer per steradian) of the zenith directly above the instrument. The SWS is a moderate resolution sensor comprised of two Zeiss spectrometers (MMS 1 NIR enhanced and NIR-PGS 2.2) for visible and near-infrared detection in the wavelength range 350 – 2170 nm. The sampling frequency is 1 Hz. The spectral resolution is 8 nm for the MMS 1 NIR and 12 nm for the NIR-PGS 2.2. The light collector is a narrow field of view (1.4°) collimator at the front end of a high-grade custom-made fiber optic bundle. The SWS does daily daytime measurements. The SWS is located in a darkroom, constructed by Southern Great Plains (SGP) site personnel within the optical trailer, to permit calibrations to be conducted without the necessity of moving the instrument to a different location. Calibrations are performed at regularly scheduled times using the ARM 12” integrating sphere.
【 预 览 】
附件列表
Files Size Format View
1020259.pdf 348KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:11次