| JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS | 卷:304 |
| A disposable glucose biosensor based on drop-coating of screen-printed carbon electrodes with magnetic nanoparticles | |
| Article | |
| Lu, Bo-Wen1  Chen, Wen-Chang1  | |
| [1] Natl Yunlin Univ Sci & Technol, Dept Chem Engn, Touliu 64002, Yunlin, Taiwan | |
| 关键词: Fe3O4 nanoparticles; Screen-printed carbon electrodes (SPCES); Electrochemical impedance spectroscopy (EIS); Glucose biosensor; | |
| DOI : 10.1016/j.jmmm.2006.01.222 | |
| 来源: Elsevier | |
PDF
|
|
【 摘 要 】
Magnetic Fe3O4 nanoparticles (Nano-Fe3O4) were prepared by co-precipitation method and a disposable glucose biosensor was fabricated by drop coating of ferricyanide (Ferri)-Nano-Fe3O4 mixture onto the surface of screen-printed carbon electrodes (SPCEs), and then by layering-on glucose oxidase (GOD). The electrochemical characteristics of modified SPCEs were analyzed by cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS) and chronoamperometry (CA). The glucose biosensors exhibit a relatively fast response (< 15 s) and high sensitivity (ca. 1.74 mu AmM-1) with a wide linear range up to 33.3mM (600mg dL(-1)) of glucose. (C) 2006 Published by Elsevier B. V.
【 授权许可】
Free
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 10_1016_j_jmmm_2006_01_222.pdf | 116KB |
PDF