【 摘 要 】
We present a portable, simple-to-operate, point-of-use analyte surface localization device. We take advantage of a set of hydrodynamic design features and components that achieve passive analyte localization by means of a single vacuum input. The vacuum source can be supplied by mechanical or battery-operated vacuum sources that are portable and allow point-of-use operation in the absence of electricity. We discuss the governing hydrodynamic principle and design parameters in detail. In a case study, we demonstrate the applicability of our technology through successful localization of a solution of rhodamine on a polydimethylsiloxane (PDMS) substrate and production of sub-millimetre-sized spots via application of a mild vacuum level of less than -10 kPa. In addition, using our devised technology, we demonstrate localized staining of human breast cancer cell blocks and human breast cancer tissue sections. (C) 2017 Elsevier B.V. All rights reserved.