Cyclic fatigue resistance of three heat-treated nickel-titanium instruments at simulated body temperature | |
article | |
Tarek M. Elsewify1  Shehabeldin M. Saber1  G Plotino1  | |
[1] Endodontic Department, Faculty of Dentistry, Ain Shams University;Department of Endodontics, The British University | |
关键词: Cyclic fatigue; Hyflex EDM; M Pro; nickel-titanium; WaveOne Gold; | |
DOI : 10.4103/sej.sej_122_19 | |
学科分类:口腔科学 | |
来源: Medknow | |
【 摘 要 】
Introduction: The aim of this study was to evaluate the cyclic fatigue resistance of three heat-treated nickel–titanium (NiTi) systems at simulated body temperature. Materials and Methods: Twelve instruments of similar apical diameter (#25) from three engine-driven NiTi instruments; WaveOne Gold (WOG), Hyflex EDM (HEDM), and M Pro system were tested for cyclic fatigue resistance at 37°C in simulated root canals with 60° angle of curvature and 5 mm curvature radius. All instruments were operated until fracture occurred. Mean and standard deviation of the time to fracture and fragment length were calculated, and statistical significance was set at 5%. The morphological characteristics of the fractured instruments were observed through scanning electron microscopy, and their chemical composition was determined using energy dispersive X-ray analysis (EDXA). Results: HEDM showed the highest resistance to cyclic fatigue at 37°C followed by WOG and M Pro instruments with a statistically significant difference. Scanning electron micrographs confirmed a predominantly ductile mode of fracture for all instruments. EDXA showed that WOG was composed of Ni (45.1 wt%) and Ti (37.0 wt%), HEDM was composed of Ni (52.9 wt%) and Ti (42.4 wt%), and M Pro was composed of Ni (49.7 wt%) and Ti (40.4 wt%). Conclusions: M Pro instrument showed the least cyclic fatigue resistance at 37°C compared to HEDM and WOG.
【 授权许可】
CC BY-NC-SA
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202307100000793ZK.pdf | 1553KB | download |