[1] Department of Biomaterials and Medical Devices Engineering, Faculty of Biomedical Engineering, Silesian University of Technology, Zabrze, Poland;Institute of Dentistry, Medical University of Warsaw, Warsaw, Poland;Institute of Engineering Materials and Biomaterials, Faculty of Mechanical Engineering, Silesian University of Technology, Gliwice, Poland | |
关键词: Keywords: Dental implant; Abutment; Zirconia ceramics; Cyclic fatigue; | |
DOI : 10.1515/amm-2016-0223 | |
学科分类:金属与冶金 | |
来源: Akademia Gorniczo-Hutnicza im. Stanislawa Staszica / University of Mining and Metallurgy | |
【 摘 要 】
The paper presents the results of finite element analysis and experimental testing under simulated physiological loading conditions on issues shaping the functional properties of individual all-ceramic abutments manufactured by CAD/CAM technology. The conducted research have cognitive significance showing the all-ceramic abutment behavior, as a key element of the implantological system, under the action of cyclic load. The aim of this study was evaluation the fatigue behavior of yttria-stabilized zirconia abutment submitted to cyclic stresses, conducted in accordance with EN ISO 14801 applies to dynamic fatigue tests of endosseous dental implants.