Strain evaluation of axially loaded collateral ligaments: a comparison of digital image correlation and strain gauges
Research
Leandra Bauer1  Manuel Kistler1  Matthias Woiczinski1  Gwendolin Prusa1  Inês Santos1  Christoph Thorwächter1 
[1] Department of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU Munich, Marchioninistraße 15, 81377, Munich, Germany;
关键词: Cadaveric biomechanical study;    Collateral ligament;    Digital image correlation;    Strain analysis;    Strain gauge;   
DOI  :  10.1186/s12938-023-01077-z
 received in 2022-08-12, accepted in 2023-02-03,  发布年份 2023
来源: Springer
PDF
【 摘 要 】

The response of soft tissue to loading can be obtained by strain assessment. Typically, strain can be measured using electrical resistance with strain gauges (SG), or optical sensors based on the digital image correlation (DIC), among others. These sensor systems are already established in other areas of technology. However, sensors have a limited range of applications in medical technology due to various challenges in handling human soft materials. The aim of this study was to compare directly attached foil-type SG and 3D-DIC to determine the strain of axially loaded human ligament structures. Therefore, the medial (MCL) and lateral (LCL) collateral ligaments of 18 human knee joints underwent cyclic displacement-controlled loading at a rate of 20 mm/min in two test trials. In the first trial, strain was recorded with the 3D-DIC system and the reference strain of the testing machine. In the second trial, strain was additionally measured with a directly attached SG. The results of the strain measurement with the 3D-DIC system did not differ significantly from the reference strain in the first trial. The strains assessed in the second trial between reference and SG, as well as between reference and 3D-DIC showed significant differences. This suggests that using an optical system based on the DIC with a given unrestricted view is an effective method to measure the superficial strain of human ligaments. In contrast, directly attached SGs provide only qualitative comparable results. Therefore, their scope on human ligaments is limited to the evaluation of changes under different conditions.

【 授权许可】

CC BY   
© The Author(s) 2023

【 预 览 】
附件列表
Files Size Format View
RO202305158436626ZK.pdf 1702KB PDF download
12302_2023_718_Article_IEq4.gif 1KB Image download
MediaObjects/12862_2023_2105_MOESM1_ESM.docx 60KB Other download
Fig. 7 673KB Image download
13046_2023_2621_Article_IEq2.gif 1KB Image download
Fig. 6 390KB Image download
MediaObjects/42004_2023_839_MOESM2_ESM.pdf 75KB PDF download
MediaObjects/41408_2023_799_MOESM2_ESM.xlsx 13KB Other download
【 图 表 】

Fig. 6

13046_2023_2621_Article_IEq2.gif

Fig. 7

12302_2023_718_Article_IEq4.gif

【 参考文献 】
  • [1]
  • [2]
  • [3]
  • [4]
  • [5]
  • [6]
  • [7]
  • [8]
  • [9]
  • [10]
  • [11]
  • [12]
  • [13]
  • [14]
  • [15]
  • [16]
  • [17]
  • [18]
  • [19]
  • [20]
  • [21]
  • [22]
  • [23]
  • [24]
  • [25]
  • [26]
  • [27]
  • [28]
  • [29]
  • [30]
  • [31]
  • [32]
  • [33]
  • [34]
  • [35]
  • [36]
  • [37]
  • [38]
  • [39]
  • [40]
  • [41]
  • [42]
  • [43]
  • [44]
  • [45]
  • [46]
  • [47]
  • [48]
  文献评价指标  
  下载次数:6次 浏览次数:1次