【 摘 要 】

Abstract Background Using a Symmetry C18 (4.6 × 150 mm, 3.5) column, a high-performance liquid chromatographic method for quantification of Rilpivirine and Cabotegravir in active pharmaceutical ingredients was developed and validated. The mobile phase is made up of buffer, acetonitrile, and 0.1 percent formic acid in a 20:80v/v ratio. The flow rate was kept constant at 1.0 ml/min, and detection was accomplished through absorption at 231 nm with a photodiode array detector. Results The calibration curve was linear, with a regression coefficient (R2) value of 0.999 and concentrations ranging from 30 to 450 g/ml of Rilpivirine and 20–300 g/ml of Cabotegravir. The method's LOD and LOQ were 0.375 g/ml, 1.238 g/ml, and 0.25 g/ml, 0.825 g/ml for Rilpivirine and Cabotegravir, respectively. Conclusions In the forced degradation studies, the degradants were characterized by using LCMS and FTIR. The current application was found to be simple, economical, and suitable, and validated according to ICH guidelines.