SAFIR-I: Design and Performance of a High-Rate Preclinical PET Insert for MRI
Jannis Fischer1  Jan Debus1  Parisa Khateri1  Volker Commichau1  Lubomir Djambazov1  Günther Dissertori1  Ulf Röser1  Christian Ritzer1  Pascal Bebié1  Mikiko Ito1  Agnieszka Zagozdzinska-Bochenek1  Robert Becker1  Werner Lustermann1  Michael Ritzert2  Peter Fischer2  Matthias T. Wyss3  Afroditi Eleftheriou3  Bruno Weber3  Geoffrey Warnock3  Charalampos Tsoumpas4 
[1] Institute for Particle Physics and Astrophysics, ETH Zürich, 8093 Zürich, Switzerland;Institute of Computer Engineering, Heidelberg University, 69120 Heidelberg, Germany;Institute of Pharmacology and Toxicology, University of Zürich, 8057 Zürich, Switzerland;Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds LS2 9JT, UK;
关键词: PET detector;    PET insert;    MRI;    preclinical PET/MRI;    dynamic PET/MR imaging;    high frame rates;   
DOI  :  10.3390/s21217037
来源: DOAJ
【 摘 要 】

(1) Background: Small Animal Fast Insert for MRI detector I (SAFIR-I) is a preclinical Positron Emission Tomography (PET) insert for the Bruker BioSpec 70/30 Ultra Shield Refrigerated (USR) preclinical 7T Magnetic Resonance Imaging (MRI) system. It is designed explicitly for high-rate kinetic studies in mice and rats with injected activities reaching 500MBq, enabling truly simultaneous quantitative PET and Magnetic Resonance (MR) imaging with time frames of a few seconds in length. (2) Methods: SAFIR-I has an axial field of view of 54.2mm and an inner diameter of 114mm. It employs Lutetium Yttrium OxyorthoSilicate (LYSO) crystals and Multi Pixel Photon Counter (MPPC) arrays. The Position-Energy-Timing Application Specific Integrated Circuit, version 6, Single Ended (PETA6SE) digitizes the MPPC signals and provides time stamps and energy information. (3) Results: SAFIR-I is MR-compatible. The system’s Coincidence Resolving Time (CRT) and energy resolution are between separate-uncertainty 209.0(3)ps and separate-uncertainty 12.41(02) Full Width at Half Maximum (FWHM) at low activity and separate-uncertainty 326.89(12)ps and separate-uncertainty 20.630(011) FWHM at 550MBq, respectively. The peak sensitivity is ∼1.6. The excellent performance facilitated the successful execution of first in vivo rat studies beyond 300MBq. Based on features visible in the acquired images, we estimate the spatial resolution to be ∼2mm in the center of the Field Of View (FOV). (4) Conclusion: The SAFIR-I PET insert provides excellent performance, permitting simultaneous in vivo small animal PET/MR image acquisitions with time frames of a few seconds in length at activities of up to 500MBq.

【 授权许可】

Unknown   

  文献评价指标  
  下载次数:0次 浏览次数:0次