期刊论文详细信息
Química Nova
Determinação de nimesulida por análise por injeção em fluxo com detecção amperométrica de múltiplos pulsos
Amanda Barbosa Lima2  Sandro Cruz Chaves2  Leonardo Morais Da Silva2  Polyana Fernandes Pereira1  Eduardo Mathias Richter1  Wallans Torres Pio Dos Santos1 
[1] ,Universidade Federal dos Vales do Jequitinhonha e Mucuri Departamento de QuímicaMG ,Brasil
关键词: pulse amperometric detection;    nimesulide;    boron-doped diamond;   
DOI  :  10.1590/S0100-40422013000900004
来源: SciELO
PDF
【 摘 要 】

A simple and fast method for the determination of nimesulide (NI) using flow injection analysis with multiple-pulse amperometric (FIA-MPA) detection at a boron-doped diamond (BDD) electrode was developed. The method was based mainly on the application of a four-potential waveform, E1(det) = -0.8 V / 30 ms, E2(det) = 0.6 V / 30 ms, E3(det) = -0.4 V / 30 ms and E4(cleaning) = -0.45 V / 300 ms versus Ag/AgCl (3.0 mol L-1 KCl). NI was detected at three different electrode potentials, at which the nitro group undergoes different redox reactions. The proposed method was selective and sensitive (detection limit of 81.0 nmol L-1), and successfully applied for the determination of NI in pharmaceutical formulations, yielding similar results to those obtained by the reference method.

【 授权许可】

CC BY-NC   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202103040064367ZK.pdf 556KB PDF download
  文献评价指标  
  下载次数:1次 浏览次数:1次