Quimica nova | |
Determination of nimesulide by flow injection analysis with multiple-pulse amperometric detection | |
Richter, Eduardo Mathias1  Pereira, Polyana Fernandes1  Silva, Leonardo Morais da2  Lima, Amanda Barbosa2  Santos, Wallans Torres Pio dos2  Chaves, Sandro Cruz2  | |
[1] Universidade Federal de Uberlândia, Brasil;Universidade Federal dos Vales do Jequitinhonha e Mucuri, Brasil | |
关键词: pulse amperometric detection; nimesulide; boron-doped diamond.; | |
DOI : 10.1590/S0100-40422013000900004 | |
学科分类:化学(综合) | |
来源: Sociedade Brasileira de Quimica | |
【 摘 要 】
A simple and fast method for the determination of nimesulide (NI) using flow injection analysis with multiple-pulse amperometric (FIA-MPA) detection at a boron-doped diamond (BDD) electrode was developed. The method was based mainly on the application of a four-potential waveform, E1(det) = -0.8 V / 30 ms, E2(det) = 0.6 V / 30 ms, E3(det) = -0.4 V / 30 ms and E4(cleaning) = -0.45 V / 300 ms versus Ag/AgCl (3.0 mol L-1 KCl). NI was detected at three different electrode potentials, at which the nitro group undergoes different redox reactions. The proposed method was selective and sensitive (detection limit of 81.0 nmol L-1), and successfully applied for the determination of NI in pharmaceutical formulations, yielding similar results to those obtained by the reference method.
【 授权许可】
CC BY
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201902015626306ZK.pdf | 556KB | download |