期刊论文详细信息
Química Nova
Determination of nimesulide by flow injection analysis with multiple-pulse amperometric detection
Silva, Leonardo Morais da1  Richter, Eduardo Mathias1  Lima, Amanda Barbosa1  Universidade Federal de Uberlândia, Brasil1  Pereira, Polyana Fernandes1  Santos, Wallans Torres Pio dos1  Chaves, Sandro Cruz1  Universidade Federal dos Vales do Jequitinhonha e Mucuri, Brasil1 
关键词: pulse amperometric detection;    nimesulide;    boron-doped diamond.;   
DOI  :  10.1590/S0100-40422013000900004
学科分类:化学(综合)
来源: Sociedade Brasileira de Quimica
PDF
【 摘 要 】

A simple and fast method for the determination of nimesulide (NI) using flow injection analysis with multiple-pulse amperometric (FIA-MPA) detection at a boron-doped diamond (BDD) electrode was developed. The method was based mainly on the application of a four-potential waveform, E1(det) = -0.8 V / 30 ms, E2(det) = 0.6 V / 30 ms, E3(det) = -0.4 V / 30 ms and E4(cleaning) = -0.45 V / 300 ms versus Ag/AgCl (3.0 mol L-1 KCl). NI was detected at three different electrode potentials, at which the nitro group undergoes different redox reactions. The proposed method was selective and sensitive (detection limit of 81.0 nmol L-1), and successfully applied for the determination of NI in pharmaceutical formulations, yielding similar results to those obtained by the reference method.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912050596114ZK.pdf 556KB PDF download
  文献评价指标  
  下载次数:2次 浏览次数:1次