Measurements of anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter using two Scheimpflug imaging devices | |
Alberto Domínguez-vicent1  Daniel Monsálvez-romín1  Antonio Jesus Del Águila-carrasco1  Santiago García-lázaro1  Robert Montés-micó1  | |
关键词: Anterior chamber/pathology; Pupil/physiology; Corneal topography; Diagnostic techniques; ophthalmological; Photography/methods; Interferometry; Câmara anterior/patologia; Pupila/fisiologia; Topografia da córnea; Técnicas de diagnóstico oftalmológico; Fotografia/métodos; Interferometria; | |
DOI : 10.5935/0004-2749.20140060 | |
来源: SciELO | |
【 摘 要 】
Purpose: To compare the ocular anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter, as measured with two different Scheimpflug imaging devices. Methods: This transversal study included 80 right eyes from 80 subjects aged from 20 to 40 years. Their spherical equivalents ranged from -4.25 to +1.00 diopters (D). Each eye's anterior chamber depth, white-to-white distance, anterior chamber angle, and pupil diameter, were measured for far vision using both the Galilei G4 (double Scheimpflug camera) and the Pentacam HR (single Scheimpflug camera) systems. Results: Mean anterior chamber depths were calculated as 3.12 ± 0.23 mm and 3.19 ± 0.24 mm when measured with the Galilei G4 and the Pentacam HR, respectively. The mean white-to-white distance measured was 11.84 ± 0.31 mm and 11.90 ± 0.43 mm when measured with the Galilei G4 and the Pentacam HR, respectively. Mean pupil diameters were measured as 3.22 ± 0.58 mm and 3.22 ± 0.52 mm when measured with the Galilei G4 and the Pentacam HR, respectively. Finally, the mean anterior chamber angle was 34.30 ± 2.86 degrees when it was measured with the Galilei G4, and 39.26 ± 2.85 degrees when measured with the Pentacam HR. A comparative analysis revealed that the Galilei G4 yielded a significantly lower (P<0.05) measurement for the anterior chamber depth, anterior chamber angle, and pupil diameter, than the Pentacam HR system. Comparable values (P>0.05) for both devices were obtained for the white-to-white distance measurements. Conclusion: The Galilei G4 and Pentacam HR Scheimpflug systems cannot be used interchangeably because they produce significant measurement differences.
【 授权许可】
CC BY
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202103040003664ZK.pdf | 331KB | download |