Repeatability of central corneal thickness measurement with the Pentacam HR system | |
Ruiz Simonato Alonso2  Bruno Machado Fontes1  Marcelo Palis Ventura2  Renato Ambrósio Jr1  | |
[1] ,Hospital Santa Casa de Misericórdia do Rio de Janeiro RJ ,Brasil | |
关键词: Cornea; Corneal topography; Biometry; Reproducibility of results; Diagnostic techniques; ophthalmological; Córnea; Topografia da córnea; Biometria; Reprodutibilidade de resultados; Técnicas de diagnostico oftamológico; | |
DOI : 10.1590/S0034-72802012000100003 | |
来源: SciELO | |
【 摘 要 】
PURPOSE: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness - CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. METHODS: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. RESULTS: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6±37 µm and 543.6±36.9 µm respectively. Mean difference between both measurements was 9.2±6.4 µm, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r² = 0,9209). CONCLUSION: Pentacam (HR) CTm gives repeatable CCT measurements over time.