| Brazilian Archives of Biology and Technology | |
| Development of an algorithm for tip-related artifacts identification in AFM biological film imaging | |
| Rubens Bernardes-filho1  Odilio Benedito Garrido De Assis1  | |
| [1] ,Embrapa Instrumentação AgropecuáriaSão Carlos SP ,Brasil | |
| 关键词: Probe microscopy; AFM; images tip artifacts; biological film; | |
| DOI : 10.1590/S1516-89132005000500020 | |
| 来源: SciELO | |
PDF
|
|
【 摘 要 】
One major drawback identified in atomic force microscopy imaging is the dependence of the image's precision on the shape of the probe tip. In this paper a simple algorithm is proposed to provide artifact identification signaling in-situ tip features in atomic force microscopy images. The base of the identifications lied when the angle formed between two scanned points was kept the same as the tip sweeps a certain length of the sample. The potential of the described method was illustrated on a chitosan polysaccharide film. The images produced were compared to evaluate tip-artifact regions. This algorithm showed promise as a tool in the measurement and characterization fields to separate true images from artificial images in probe microscopy.
【 授权许可】
CC BY-NC
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO202005130165364ZK.pdf | 173KB |
PDF