期刊论文详细信息
Journal of the Brazilian Chemical Society
Visible LED-Based Instrumentation for Photometric Determination of Electroosmotic Flow in Microchannels
José Alberto Fracassi Da Silva2  Ariane Deblire2  Dosil P. De Jesus2  Wendell Karlos T. Coltro1 
[1] ,Universidade Estadual de Campinas Instituto de Química Campinas SP ,Brazil
关键词: electroosmotic flow;    μTAS;    Lab-on-a-chip;    photometric methods;    microfluidic devices;   
DOI  :  10.1590/S0103-50532011000400017
来源: SciELO
PDF
【 摘 要 】

An instrumental setup for electroosmotic flow (EOF) determination in microfluidic devices is described. The system is based on a trinocular microscope and an integrated photodiode OPT101. The radiation was provided by a high-intensity white LED. For the determinations, patent blue V and methylene blue aqueous solutions were used. The setup was applied for EOF monitoring in hybrid toner/glass and PDMS microchips. For the toner/glass device, the EOF mobility was determined to be (5.75 ± 0.01) 10-4 cm² V-1 s-1 at pH 7.0. For PDMS devices the measured EOF mobility was (3.2 ± 0.1) 10-4 cm² V-1 s-1. Reproducible measurements were obtained in all experiments, which produced results with small errors. The proposed method was compared to the conventional current monitoring method.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130106472ZK.pdf 1542KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:11次