期刊论文详细信息
Química Nova
Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants
Ana Gabriela Reis Solano2  Gisele Rodrigues Da Silva1  Silvia Ligório Fialho1  Armando Da Silva Cunha Júnior1  Gérson Antônio Pianetti1 
[1] ,Universidade Federal de São João Del Rei Faculdade de Farmácia Divinópolis MG ,Brasil
关键词: validation;    etoposide;    poly(ε-caprolactone);    intraocular implant;   
DOI  :  10.1590/S0100-40422012000600032
来源: SciELO
PDF
【 摘 要 】

A method using HPLC-UV was developed and validated for the determination of etoposide incorporated into polycaprolactone implants. The method was carried out in isocratic mode using a C18 column (250 x 4.6 mm; 5 µm), at 25 ºC, with acetonitrile and acetic acid 4% (70:30) as mobile phase, a flow rate of 2 mL/min, and UV detection at 285 nm. The method was linear (r² > 0.99) over the range of 5 to 65 µg/mL, precise (RSD < 5%), accurate (recovery of 98.7%), robust, selective regarding excipient of the sample, and had a quantitation limit equal to 1.76 µg/mL. The validated method can be successfully employed for routine quality control analyses.