Quimica nova | |
Development and validation of a High Performance Liquid Chromatographic method for determination of etoposide in biodegradable polymeric implants | |
Fialho, Silvia Ligório1  Pianetti, Gérson Antônio2  Cunha Júnior, Armando da Silva2  Silva, Gisele Rodrigues da3  Solano, Ana Gabriela Reis3  | |
[1] Fundação Ezequiel Dias, Belo Horizonte, Brasil;Universidade Federal de Minas Gerais, Belo Horizonte, Brasil;Universidade Federal de São João Del Rei, Divinópolis, Brasil | |
关键词: validation; etoposide; poly(ε; -; caprolactone); intraocular implant.; | |
DOI : 10.1590/S0100-40422012000600032 | |
学科分类:化学(综合) | |
来源: Sociedade Brasileira de Quimica | |
【 摘 要 】
A method using HPLC-UV was developed and validated for the determination of etoposide incorporated into polycaprolactone implants. The method was carried out in isocratic mode using a C18 column (250 x 4.6 mm; 5 µm), at 25 ºC, with acetonitrile and acetic acid 4% (70:30) as mobile phase, a flow rate of 2 mL/min, and UV detection at 285 nm. The method was linear (r2 > 0.99) over the range of 5 to 65 µg/mL, precise (RSD < 5%), accurate (recovery of 98.7%), robust, selective regarding excipient of the sample, and had a quantitation limit equal to 1.76 µg/mL. The validated method can be successfully employed for routine quality control analyses.
【 授权许可】
CC BY
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201902014466879ZK.pdf | 485KB | download |