Silva, Gisele Rodrigues da1  Universidade Federal de São João Del Rei, Divinópolis, Brasil1  Fialho, Silvia Ligório1  Fundação Ezequiel Dias, Belo Horizonte, Brasil1  Solano, Ana Gabriela Reis1  Universidade Federal de Minas Gerais, Belo Horizonte, Brasil1  Pianetti, Gérson Antônio1  Cunha Júnior, Armando da Silva1 
关键词: validation;    etoposide;    poly(ε;    -;    caprolactone);    intraocular implant.;   
DOI  :  10.1590/S0100-40422012000600032
学科分类:化学(综合)
来源: Sociedade Brasileira de Quimica
PDF
【 摘 要 】
A method using HPLC-UV was developed and validated for the determination of etoposide incorporated into polycaprolactone implants. The method was carried out in isocratic mode using a C18 column (250 x 4.6 mm; 5 µm), at 25 ºC, with acetonitrile and acetic acid 4% (70:30) as mobile phase, a flow rate of 2 mL/min, and UV detection at 285 nm. The method was linear (r2 > 0.99) over the range of 5 to 65 µg/mL, precise (RSD < 5%), accurate (recovery of 98.7%), robust, selective regarding excipient of the sample, and had a quantitation limit equal to 1.76 µg/mL. The validated method can be successfully employed for routine quality control analyses.