期刊论文详细信息
Química Nova
Design and evaluation of a continuous flow, integrated nebulizer-hydride generator for flame atomic absorption spectrometry
Miguel Murillo1  Nereida Carrión1  Jorge Colmenares1  Juan Romero1  Gladys Alvarado1  Manuel Ríos1  Freddy Angulo1 
[1] ,Universidad Central de Venezuela Facultad de Ciencias Escuela de QuímicaCaracas,Venezuela
关键词: nebulizer-hydride generation;    arsenic;    selenium;   
DOI  :  10.1590/S0100-40422008000600008
来源: SciELO
PDF
【 摘 要 】

An evaluation of the performance of a continuous flow hydride generator-nebulizer for flame atomic absorption spectrometry was carried out. Optimization of nebulizer gas flow rate, sample acid concentration, sample and tetrahydroborate uptake rates and reductant concentration, on the As and Se absorbance signals was carried out. A hydrogen-argon flame was used. An improvement of the analytical sensitivity relative to the conventional bead nebulizer used in flame AA was obtained (2 (As) and 4.8 (Se) µg L-1). Detection limits (3σb) of 1 (As) and 1.3 (Se) µg L-1 were obtained. Accuracy of the method was checked by analyzing an oyster tissue reference material.

【 授权许可】

CC BY-NC   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130062508ZK.pdf 175KB PDF download
  文献评价指标  
  下载次数:0次 浏览次数:0次