Revista Brasileira de Oftalmologia
Repeatability of central corneal thickness measurement with the Pentacam HR system
Ruiz Simonato Alonso2  Bruno Machado Fontes1  Marcelo Palis Ventura2  Renato Ambrósio Jr1 
[1] ,Hospital Santa Casa de Misericórdia do Rio de Janeiro RJ ,Brasil
关键词: Cornea;    Corneal topography;    Biometry;    Reproducibility of results;    Diagnostic techniques;    ophthalmological;    Córnea;    Topografia da córnea;    Biometria;    Reprodutibilidade de resultados;    Técnicas de diagnostico oftamológico;   
DOI  :  10.1590/S0034-72802012000100003
来源: SciELO
PDF
【 摘 要 】

PURPOSE: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness - CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. METHODS: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. RESULTS: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6±37 µm and 543.6±36.9 µm respectively. Mean difference between both measurements was 9.2±6.4 µm, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r² = 0,9209). CONCLUSION: Pentacam (HR) CTm gives repeatable CCT measurements over time.

【 授权许可】

CC BY   
 All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License

【 预 览 】
附件列表
Files Size Format View
RO202005130022687ZK.pdf 446KB PDF download
  文献评价指标  
  下载次数:10次 浏览次数:18次