Arquivos de Neuro-Psiquiatria | |
Portable lactate analyzer for measuring lactate in cerebrospinal fluid (CSF) and plasma ? method-comparison evaluations | |
Sérgio Monteiro De Almeida1  Jennifer Marquie-beck1  Archana Bhatt1  Scott Letendre1  Allen Mccutchan1  Ron Ellis1  | |
关键词: HIV; cerebrospinal fluid; clinical chemistry; lactate; HIV; líquido cefalorraquidiano; química clínica; lactato; | |
DOI : 10.1590/0004-282X20140076 | |
来源: SciELO | |
【 摘 要 】
Increased plasma lactate levels can indicate the presence of metabolic disorders in HIV infected individuals.Objective:To determine whether a portable analyzer is valid for measuring cerebrospinal fluid (CSF) and plasma lactate levels in HIV infected individuals. Method:CSF and plasma were collected from 178 subjects. Samples tested by the Accutrend® portable analyzer were compared to those tested by a reference device (SYNCHRON LX® 20). Results:The portable analyzer had in plasma sensitivity of 0.95 and specificity 0.87. For CSF the specificity was 0.95; the sensitivity 0.33; the negative predictive value was 95% and the positive predictive value 33%. Conclusions:These findings support the validity of the portable analyzer in measuring lactate concentrations in CSF that fall within the normal range. The relatively poor positive predictive value indicates that a result above the reference range may represent a “false positive test”, and should be confirmed by the reference device before concluding abnormality.
【 授权许可】
CC BY
All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License