期刊论文详细信息
Sensors
Effect of Silver Annealing Conditions on the Performance of Electrolytic Silver/Silver Chloride Electrodes used in Harned Cell Measurements of pH
Paul. J. Brewer1 
关键词: pH;    Harned cell;    Ag/AgCl electrodes;    annealing;    stability;   
DOI  :  10.3390/s100302202
来源: mdpi
PDF
【 摘 要 】

We have studied the long and short term stability of electrolytic Ag/AgCl electrodes fabricated from Ag wire that has been subjected to a range of different annealing conditions. At elevated temperatures, the presence of oxygen during the annealing process has been shown to be detrimental to the performance of electrodes produced. This phenomenon has been attributed to the dissolution of oxygen in the Ag lattice leading to structural changes in the Ag/AgCl electrode material. Electrodes prepared from Ag wire annealed in the absence of oxygen have shown no appreciable change in performance throughout the temperature range employed. This work has resulted in an improved understanding of the optimum annealing conditions required for Ag used in the preparation of electrolytic Ag/AgCl reference electrodes. This work has positive implications for the accuracy of Harned cell measurements of pH.

【 授权许可】

CC BY   
© 2010 by the authors; licensee Molecular Diversity Preservation International, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190054543ZK.pdf 2645KB PDF download
  文献评价指标  
  下载次数:16次 浏览次数:18次