会议论文详细信息
3rd International Conference on New Material and Chemical Industry
Stability study of Pt as reference electrode for ISFETs operating in reference electrolyte
材料科学;化学工业
Zhang, J.K.^1 ; Zhao, D.^1 ; Yang, H.^1 ; Li, C.^1 ; Xu, M.^1 ; Wu, D.P.^1
State Key Laboratory of ASIC and System, Fudan University, Shanghai
200433, China^1
关键词: Ag/AgCl electrodes;    Bio-chemical sensor;    Compositional changes;    Good stability;    Pt electrode;    Reference electrodes;    Stability study;   
Others  :  https://iopscience.iop.org/article/10.1088/1757-899X/479/1/012042/pdf
DOI  :  10.1088/1757-899X/479/1/012042
学科分类:材料科学(综合)
来源: IOP
PDF
【 摘 要 】

Pt electrode used as reference electrode (RE) of Ion Sensitive Field-Effect Transistor (ISFET) is controversial with respect to its redox reactions with the electrolyte so that its potential can be easily perturbed by compositional changes or the flow of electrolyte. The microfluidic-controlled Pt-KCl RE system (PtRE) has been previously proposed to potentially replace the conventional RE for specific applications. This paper presents our further study on the stability of Pt electrode as ISFET's RE using the PtRE system. It is found that the potential of Pt electrode preserved in air is not stable in the beginning and drifts slowly for quite a few minutes. However, by immersing the Pt electrode in KCl reference electrolyte for a certain time prior to test, the reference characteristics of the Pt electrode can be greatly improved and exhibit as almost good stability performance as conventional Ag/AgCl electrode. Hence, Pt electrode in the PtRE system can be used as a viable alternative to conventional RE for ISFETs and other biochemical sensors.

【 预 览 】
附件列表
Files Size Format View
Stability study of Pt as reference electrode for ISFETs operating in reference electrolyte 769KB PDF download
  文献评价指标  
  下载次数:36次 浏览次数:31次