Sensors | |
Error Analysis and Measurement Uncertainty for a Fiber Grating Strain-Temperature Sensor | |
Jaw-Luen Tang1  | |
[1] Department of Physics, National Chung Cheng University, 168 University Road, Chia-Yi 62102, Taiwan; E-Mail: | |
关键词: fiber Bragg grating; sensor; strain; temperature; error analysis; measurement uncertainty; | |
DOI : 10.3390/s100706582 | |
来源: mdpi | |
【 摘 要 】
A fiber grating sensor capable of distinguishing between temperature and strain, using a reference and a dual-wavelength fiber Bragg grating, is presented. Error analysis and measurement uncertainty for this sensor are studied theoretically and experimentally. The measured root mean squared errors for temperature T and strain ε were estimated to be 0.13 °C and 6 με, respectively. The maximum errors for temperature and strain were calculated as 0.00155 T + 2.90 × 10−6 ε and 3.59 × 10−5 ε + 0.01887 T, respectively. Using the estimation of expanded uncertainty at 95% confidence level with a coverage factor of
【 授权许可】
CC BY
© 2010 by the authors licensee MDPI, Basel, Switzerland.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202003190053136ZK.pdf | 676KB | download |