期刊论文详细信息
Sensors
Error Analysis and Measurement Uncertainty for a Fiber Grating Strain-Temperature Sensor
Jaw-Luen Tang1 
[1] Department of Physics, National Chung Cheng University, 168 University Road, Chia-Yi 62102, Taiwan; E-Mail:
关键词: fiber Bragg grating;    sensor;    strain;    temperature;    error analysis;    measurement uncertainty;   
DOI  :  10.3390/s100706582
来源: mdpi
PDF
【 摘 要 】

A fiber grating sensor capable of distinguishing between temperature and strain, using a reference and a dual-wavelength fiber Bragg grating, is presented. Error analysis and measurement uncertainty for this sensor are studied theoretically and experimentally. The measured root mean squared errors for temperature T and strain ε were estimated to be 0.13 °C and 6 με, respectively. The maximum errors for temperature and strain were calculated as 0.00155 T + 2.90 × 10−6 ε and 3.59 × 10−5 ε + 0.01887 T, respectively. Using the estimation of expanded uncertainty at 95% confidence level with a coverage factor of k = 2.205, temperature and strain measurement uncertainties were evaluated as 2.60 °C and 32.05 με, respectively. For the first time, to our knowledge, we have demonstrated the feasibility of estimating the measurement uncertainty for simultaneous strain-temperature sensing with such a fiber grating sensor.

【 授权许可】

CC BY   
© 2010 by the authors licensee MDPI, Basel, Switzerland.

【 预 览 】
附件列表
Files Size Format View
RO202003190053136ZK.pdf 676KB PDF download