Sensors | |
Development of Robust and Standardized Cantilever Sensors Based on Biotin/Neutravidin Coupling for Antibody Detection | |
Jiayun Zhang2  Hans Peter Lang2  Felice Battiston1  Natalija Backmann2  Francois Huber2  | |
[1] Concentris GmbH, Davidsbodenstrasse 63, 4056 Basel, Switzerland; E-Mail:;Swiss Nano Institute, University of Basel, Klingelbergstrasse 82, 4056 Basel, Switzerland; E-Mails: | |
关键词: cantilever; biosensor; protein; multilayer; NeutrAvidin; biotin; | |
DOI : 10.3390/s130405273 | |
来源: mdpi | |
【 摘 要 】
A cantilever-based protein biosensor has been developed providing a customizable multilayer platform for the detection of antibodies. It consists of a biotin-terminated PEG layer pre-functionalized on the gold-coated cantilever surface, onto which NeutrAvidin is adsorbed through biotin/NeutrAvidin specific binding. NeutrAvidin is used as a bridge layer between the biotin-coated surface and the biotinylated biomolecules, such as biotinylated bovine serum albumin (biotinylated BSA), forming a multilayer sensor for direct antibody capture. The cantilever biosensor has been successfully applied to the detection of mouse anti-BSA (m-IgG) and sheep anti-BSA(s-IgG) antibodies. As expected, the average differential surface stress signals of about 5.7 ± 0.8 × 10−3 N/m are very similar for BSA/m-IgG and BSA/s-IgG binding,
【 授权许可】
CC BY
© 2013 by the authors; licensee MDPI, Basel, Switzerland.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO202003190036901ZK.pdf | 955KB | download |