期刊论文详细信息
Metrology and Measurement Systems
Measuring the Stray Capacitance of Solenoids with a Transmitting and a Receiving Coil
A. Mariscotti1 
关键词: Capacitance measurement;    Uncertainty;    Stray capacitance;    Linear regression;    Resonance condition;   
DOI  :  10.2478/v10178-011-0005-4
来源: Versita
PDF
【 摘 要 】

Air core solenoids, possibly single layer and with significant spacing between turns, are often used to ensure low stray capacitance, as they are used as part of many sensors and instruments. The problem of the correct estimation of the stray capacitance is relevant both during design and to validate measurement results; the expected value is so low to be influenced by any stray capacitance of the external measurement instrument. A simplified method is proposed that does not perturb the stray capacitance of the solenoid under test; the method is based on resonance with an external capacitor and on the use of a linear regression technique.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912080706795ZK.pdf 194KB PDF download
  文献评价指标  