| Journal of the Brazilian Chemical Society | |
| Determination of metoprolol, acebutolol and propranolol in pharmaceutical formulations using the same SIA system | |
| Korn, M. Graças A.1  Saraiva, M. Lúcia M. F. S.1  Passos, Marieta L. C.1  Universidade Federal da Bahia, Salvador, Brazil1  Lima, José L. F. C.1  Universidade do Porto, Porto, Portugal1  | |
| 关键词: beta-blockers; sequential injection analysis; SIA; spectrophotometry; pharmaceutical formulations; | |
| DOI : 10.1590/S0103-50532008000300027 | |
| 学科分类:化学(综合) | |
| 来源: SciELO | |
PDF
|
|
【 摘 要 】
A mechanized system for the determination of bblockers in pharmaceutical formulations is presented. Using the Sequential Injection Analysis (SIA) technique it was achieved a simple, economical and versatile methodology adaptable to any pharmaceutical control involving these substances. It does not require any pre-treatment for the samples, as they must only be dissolved in acid before analysis. Linear calibration plots were obtained for metoprolol (40.52 - 250 mg L-1), acebutolol (32.85 - 140 mg L-1) and propranolol (16.58 - 120 mg L-1). A R.S.D. lower than 5% was attained. The methodology was used in tablets, injections and prolonged-release capsules. Common excipients used in pharmaceuticals do not interfere. Statistical comparison of the results obtained with the proposed methodology and with the official methods showed good agreement (95% confidence level). SIA system produces only 2.50 mL of effluents per determination whereas the reference methodologies consume around 140 mL.
【 授权许可】
Unknown
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| RO201912050580395ZK.pdf | 140KB |
PDF