期刊论文详细信息
Journal of Chemical Sciences
An indigenous cluster beam apparatus with a reflectron time-of-flight mass spectrometer
C N R Rao2  G U Kulkarni2  G Raina2  R T Yadav2  V S Ramamurthy1 
[1] Department of Science and Technology, Technology Bhavan, New Mehrauli Road, New Delhi 110 016, India$$;Chemistry and Physics of Materials Unit, Jawaharlal Nehru Centre for Advanced Scientific Research, Jakkur PO, Bangalore 560 064, India$$
关键词: Mass spectrometer;    cluster source;    instrumentation.;   
DOI  :  
来源: Indian Academy of Sciences
PDF
【 摘 要 】

The design and fabrication of a Smalley-type cluster source in combination with a reflectron based time-of-flight (TOF) mass spectrometer are reported. The generation of clusters is based on supersonic jet expansion of the sampling plume. Sample cells for both liquid and solid targets developed for this purpose are described. Two pulsed Nd-YAG lasers are used in tandem, one (532 nm) for target vapourization and the other (355 nm) for cluster ionization. Methanol clusters of nuclearity up to 14 (mass 500 amu) were produced from liquid methanol as the test sample. The clusters were detected with a mass resolution of ~ 2500 in the R-TOF geometry. Carbon clusters up to a nuclearity of 28 were obtained using a polyimide target. The utility of the instrument is demonstrated by carrying out experiments to generate mixed clusters from alcohol mixtures.

【 授权许可】

Unknown   

【 预 览 】
附件列表
Files Size Format View
RO201912040507241ZK.pdf 339KB PDF download
  文献评价指标  
  下载次数:5次 浏览次数:10次