Journal of Nuclear Medicine Technology | |
18-Month Performance Assessment of Gemini TF 16 PET/CT System in a High-Volume Department | |
Ashish Kumar Jha1  Abhijith Mohan Singh1  Venkatesh Rangarajan1  Nilendu C. Purandare1  Sneha Mithun1  Archi Agrawal1  Sneha Shah1  | |
[1] Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital, Parel, Mumbai, India Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital, Parel, Mumbai, India Department of Nuclear Medicine and Molecular Imaging, Tata Memorial Hospital, Parel, Mumbai, India | |
关键词: PET; PET/CT; time-of-flight; fully 3-dimensional whole-body imaging; LSO; LYSO; NEMA QC; | |
DOI : 10.2967/jnmt.115.168492 | |
学科分类:医学(综合) | |
来源: Society of Nuclear Medicine | |
【 摘 要 】
Acceptance testing is a set of quality control tests performed to verify various manufacturer-specified parameters before a newly installed PET/CT system can be accepted for clinical use. A new PET/CT system, Gemini TF 16, installed in our department in September 2012 has a PET component capable of time-of-flight imaging using lutetium-yttrium-oxyorthosilicate crystals and operates in 3-dimensional mode. Our aim was to evaluate the system before acceptance and observe the consistency of its performance during high-volume work for 18 mo after installation (we perform an average of 30 PET/CT scans daily). Methods: We performed NEMA (National Electrical Manufacturers Association) NU-2 2007 acceptance testing on the Gemini TF 16; continuously evaluated its gain calibration, timing resolution, and energy resolution during the subsequent 18 mo; and analyzed the results. Results: The system passed the acceptance testing and showed few fluctuations in energy and timing resolutions during the observation period. Conclusion: The Gemini TF 16 whole-body PET/CT system performed excellently during the 18-mo study period despite the high volume of work.
【 授权许可】
Unknown
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RO201912010179944ZK.pdf | 428KB | download |