Fast determination of basic drugs in plasma by gas chromatography with nitrogen phosphorus detector
Lanaro, Rafael1  Linden, Rafael2  Silva, Luana Letícia da2  Lizot, Lilian Feltraco2  Antunes, Marina Venzon2  Spaniol, Bárbara2 
[1] Universidade Estadual de Campinas, Campinas, Brasil;Universidade Feevale, Novo Hamburgo, Brasil
关键词: GC-;    NPD;    emergency toxicology;    drug screening.;   
DOI  :  10.1590/S0100-40422012000600029
学科分类:化学(综合)
来源: Sociedade Brasileira de Quimica
PDF
【 摘 要 】

A simple and fast method for determination of 40 basic drugs in human plasma employing gas-chromatography with nitrogen-phosphorus detection was developed and validated. Drugs were extracted from 800 µL of plasma with 250 µL of butyl acetate at basic pH. Aliquots of the organic extract were directly injected on a column with methylsilicone stationary phase. Total chromatographic run time was 25 min. All compounds were detected in concentrations ranging from therapeutic to toxic levels, with intermediate precision CV% below 11.2 and accuracy in the range of 92-114%.