Scandinavian Journal of Trauma, Resuscitation and Emergency Medicine | |
Effect of regulating airway pressure on intrathoracic pressure and vital organ perfusion pressure during cardiopulmonary resuscitation: a non-randomized interventional cross-over study | |
Keith Lurie1  Demetris Yannopoulos2  Scott McKnite1  Jennifer Rees1  Anja Metzger1  Laura Puertas1  Guillaume Debaty1  Younghoon Kwon3  | |
[1] Department of Emergency Medicine, University of Minnesota, CMinneapolis, MN, USA;Department of Medicine-Cardiovascular Division, University of Minnesota, Minneapolis, MN, USA;Department of Medicine-Cardiovascular Division, University of Virginia, Charlottesville, VA, USA | |
关键词: Intrathoracic pressure regulation; Intrapleural pressure; Impedance threshold device; Cardiopulmonary resuscitation; Airway pressure; | |
Others : 1231197 DOI : 10.1186/s13049-015-0164-5 |
|
received in 2015-05-29, accepted in 2015-10-20, 发布年份 2015 | |
【 摘 要 】
Background
The objective of this investigation was to evaluate changes in intrathoracic pressure (Ppl), airway pressure (Paw) and vital organ perfusion pressures during standard and intrathoracic pressure regulation (IPR)-assisted cardiopulmonary resuscitation (CPR).
Methods
Multiple CPR interventions were assessed, including newer ones based upon IPR, a therapy that enhances negative intrathoracic pressure after each positive pressure breath. Eight anesthetized pigs underwent 4 min of untreated ventricular fibrillation followed by 2 min each of sequential interventions: (1) conventional standard CPR (STD), (2) automated active compression decompression (ACD) CPR, (3) ACD+ an impedance threshold device (ITD) CPR or (4) ACD+ an intrathoracic pressure regulator (ITPR) CPR, the latter two representing IPR-based CPR therapies. Intrapleural (Ppl), airway (Paw), right atrial, intracranial, and aortic pressures, along with carotid blood flow and end tidal CO 2 , were measured and compared during each CPR intervention.
Results
The lowest mean and decompression phase Ppl were observed with IPR-based therapies [Ppl mean (mean ± SE): STD (0.8 ± 1.1 mmHg); ACD (−1.6 ± 1.6); ACD-ITD (−3.7 ± 1.5, p < 0.05 vs. both STD and ACD); ACD-ITPR (−7.0 ± 1.9, p < 0.05 vs. both STD and ACD)] [Ppl decompression (mean ± SE): STD (−6.3 ± 2.2); ACD (−13.0 ± 3.8); ACD-ITD −16.9 ± 3.6, p < 0.05 vs. both STD and ACD); ACD-ITPR −18.7 ± 3.5, p < 0.05 vs. both STD and ACD)]. Interventions with the lower mean or decompression phase Ppl also demonstrated lower Paw and were associated with higher vital organ perfusion pressures.
Conclusions
IPR-based CPR methods, specifically ACD-ITPR, yielded the most pronounced reduction in both Ppl and Paw and resulted in the most favorable augmentation of hemodynamics during CPR.
【 授权许可】
2015 Kwon et al.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20151109082435666.pdf | 1463KB | download | |
Fig. 5. | 39KB | Image | download |
Fig. 4. | 33KB | Image | download |
Fig. 3. | 31KB | Image | download |
Fig. 2. | 76KB | Image | download |
Fig. 1. | 8KB | Image | download |
【 图 表 】
Fig. 1.
Fig. 2.
Fig. 3.
Fig. 4.
Fig. 5.
【 参考文献 】
- [1]Rudikoff MT, Maughan WL, Effron M, Freund P, Weisfeldt ML. Mechanisms of blood flow during cardiopulmonary resuscitation. Circulation. 1980; 61(2):345-352.
- [2]Kouwenhoven WB, Jude JR, Knickerbocker GG. Closed-chest cardiac massage. Jama. 1960; 173:1064-1067.
- [3]Lurie KG, Lindo C, Chin J. CPR: the P stands for plumber’s helper. JAMA. 1990; 264(13):1661.
- [4]Perkins GD, Brace S, Gates S. Mechanical chest-compression devices: current and future roles. Curr Opin Crit Care. 2010; 16(3):203-210.
- [5]Aufderheide TP, Frascone RJ, Wayne MA, Mahoney BD, Swor RA, Domeier RM et al.. Standard cardiopulmonary resuscitation versus active compression-decompression cardiopulmonary resuscitation with augmentation of negative intrathoracic pressure for out-of-hospital cardiac arrest: a randomised trial. Lancet. 2011; 377(9762):301-311.
- [6]Ma MH, Hwang JJ, Lai LP, Wang SM, Huang GT, Shyu KG et al.. Transesophageal echocardiographic assessment of mitral valve position and pulmonary venous flow during cardiopulmonary resuscitation in humans. Circulation. 1995; 92(4):854-861.
- [7]Niemann JT, Rosborough JP, Hausknecht M, Garner D, Criley JM. Pressure-synchronized cineangiography during experimental cardiopulmonary resuscitation. Circulation. 1981; 64(5):985-991.
- [8]Weisfeldt ML, Chandra N. Physiology of cardiopulmonary resuscitation. Annu Rev Med. 1981; 32:435-442.
- [9]Sigurdsson G, Yannopoulos D, McKnite SH, Lurie KG. Cardiorespiratory interactions and blood flow generation during cardiac arrest and other states of low blood flow. Curr Opin Crit Care. 2003; 9(3):183-188.
- [10]Sasson C, Meischke H, Abella BS, Berg RA, Bobrow BJ, Chan PS et al.. Increasing cardiopulmonary resuscitation provision in communities with low bystander cardiopulmonary resuscitation rates: a science advisory from the American Heart Association for healthcare providers, policymakers, public health departments, and community leaders. Circulation. 2013; 127(12):1342-1350.
- [11]Lurie KG, Coffeen P, Shultz J, McKnite S, Detloff B, Mulligan K. Improving active compression-decompression cardiopulmonary resuscitation with an inspiratory impedance valve. Circulation. 1995; 91(6):1629-1632.
- [12]Yannopoulos D, Nadkarni VM, McKnite SH, Rao A, Kruger K, Metzger A et al.. Intrathoracic pressure regulator during continuous-chest-compression advanced cardiac resuscitation improves vital organ perfusion pressures in a porcine model of cardiac arrest. Circulation. 2005; 112(6):803-811.
- [13]Shultz JJ, Coffeen P, Sweeney M, Detloff B, Kehler C, Pineda E et al.. Evaluation of standard and active compression-decompression CPR in an acute human model of ventricular fibrillation. Circulation. 1994; 89(2):684-693.
- [14]Turner JM, Mead J, Wohl ME. Elasticity of human lungs in relation to age. J Appl Physiol. 1968; 25(6):664-671.
- [15]Lurie KG, Yannopoulos D, McKnite SH, Herman ML, Idris AH, Nadkarni VM et al.. Comparison of a 10-breaths-per-minute versus a 2-breaths-per-minute strategy during cardiopulmonary resuscitation in a porcine model of cardiac arrest. Respir Care. 2008; 53(7):862-870.
- [16]Segal N, Parquette B, Ziehr J, Yannopoulos D, Lindstrom D. Intrathoracic pressure regulation during cardiopulmonary resuscitation: a feasibility case-series. Resuscitation. 2013; 84(4):450-453.
- [17]Yannopoulos D, Aufderheide TP, McKnite S, Kotsifas K, Charris R, Nadkarni V et al.. Hemodynamic and respiratory effects of negative tracheal pressure during CPR in pigs. Resuscitation. 2006; 69(3):487-494.
- [18]Hausknecht MJ, Wise RA, Brower RG, Hassapoyannes C, Weisfeldt ML, Suzuki J et al.. Effects of lung inflation on blood flow during cardiopulmonary resuscitation in the canine isolated heart-lung preparation. Circ Res. 1986; 59(6):676-683.
- [19]Chandra N, Guerci A, Weisfeldt ML, Tsitlik J, Lepor N. Contrasts between intrathoracic pressures during external chest compression and cardiac massage. Crit Care Med. 1981; 9(11):789-792.
- [20]Metzger AK, Herman M, McKnite S, Tang W, Yannopoulos D. Improved cerebral perfusion pressures and 24-hr neurological survival in a porcine model of cardiac arrest with active compression-decompression cardiopulmonary resuscitation and augmentation of negative intrathoracic pressure. Crit Care Med. 2012; 40(6):1851-1856.
- [21]Yannopoulos D, Metzger A, McKnite S, Nadkarni V, Aufderheide TP, Idris A et al.. Intrathoracic pressure regulation improves vital organ perfusion pressures in normovolemic and hypovolemic pigs. Resuscitation. 2006; 70(3):445-453.
- [22]Aufderheide TP, Lurie KG. Vital organ blood flow with the impedance threshold device. Crit Care Med. 2006; 34(12 Suppl):S466-S473.
- [23]Lurie KG, Mulligan KA, McKnite S, Detloff B, Lindstrom P, Lindner KH. Optimizing standard cardiopulmonary resuscitation with an inspiratory impedance threshold valve. Chest. 1998; 113(4):1084-1090.
- [24]Wenzel V, Fuerst RS, Idris AH, Banner MJ, Rush WJ, Orban DJ. Automatic mechanical device to standardize active compression-decompression CPR. Ann Emerg Med. 1995; 25(3):386-389.
- [25]Chang MW, Coffeen P, Lurie KG, Shultz J, Bache RJ, White CW. Active compression-decompression CPR improves vital organ perfusion in a dog model of ventricular fibrillation. Chest. 1994; 106(4):1250-1259.
- [26]Lurie KG, Zielinski T, McKnite S, Aufderheide T, Voelckel W. Use of an inspiratory impedance valve improves neurologically intact survival in a porcine model of ventricular fibrillation. Circulation. 2002; 105(1):124-129.
- [27]Lurie KG, Voelckel WG, Zielinski T, McKnite S, Lindstrom P, Peterson C et al.. Improving standard cardiopulmonary resuscitation with an inspiratory impedance threshold valve in a porcine model of cardiac arrest. Anesth Analg. 2001; 93(3):649-655.
- [28]Plaisance P, Lurie KG, Payen D. Inspiratory impedance during active compression-decompression cardiopulmonary resuscitation: a randomized evaluation in patients in cardiac arrest. Circulation. 2000; 101(9):989-994.
- [29]Simmons DH, Linde LM, Miller JH, O’Reilly RJ. Relation between lung volume and pulmonary vascular resistance. Circ Res. 1961; 9(2):465-471.
- [30]Yannopoulos D, Zviman M, Castro V, Kolandaivelu A, Ranjan R, Wilson RF et al.. Intra-cardiopulmonary resuscitation hypothermia with and without volume loading in an ischemic model of cardiac arrest. Circulation. 2009; 120(14):1426-1435.
- [31]Ditchey RV, Lindenfeld J. Potential adverse effects of volume loading on perfusion of vital organs during closed-chest resuscitation. Circulation. 1984; 69(1):181-189.
- [32]Lurie K, Zielinski T, McKnite S, Sukhum P. Improving the efficiency of cardiopulmonary resuscitation with an inspiratory impedance threshold valve. Crit Care Med. 2000; 28(11 Suppl):N207-N209.