Radiographic and safety details of vertebral body stenting: results from a multicenter chart review | |
Paul Friedhelm Heini2  Lorin Michael Benneker7  Robert Rotter3  Gianluca Maestretti5  Soren Eiskjaer4  Sebastian Fürderer1  Frank Kandziora6  Matti Scholz6  Thomas Vordemvenne8  Gosia Perler9  Christoph Röder9  Peter Diel6  | |
[1] Department of Spine Surgery, Mutterhaus der Borromäerinnen Hospital Trier, Feldstraße 16, 54290 Trier, Germany;Spine Center, Sonnenhof Hospital, Buchserstrasse 30, 3006 Bern, Switzerland;Department of Spine Surgery, University Hospital Rostock, Schillingallee 35, 18057 Rostock, Germany;Department of Spine Surgery, Aalborg University Hospital, Hobrovej18–22, 9100, Aalborg, Denmark;Department of Spine Surgery, Cantonal Hospital Fribourg, Chemin des Pensionnats 2, 1708 Fribourg, Switzerland;Center for Spinal Surgery and Neurotraumatology, BGU Hospital Frankfurt, Friedberger Landstraße 430, 60389 Frankfurt, Germany;Department of Orthopaedic Surgery, Inselspital, University Hospital Bern, Freiburgstrasse, 3010 Bern, Switzerland;Department of Trauma Surgery, University Hospital Münster, Domagkstraße 5, 48149 Münster, Germany;MEM Research Center, Institute for Evaluative Research in Orthopaedic Surgery, University of Bern, Stauffacherstrasse 78, CH-3014 Bern, Switzerland | |
关键词: Osteoporosis; Stentoplasty; Vertebral body stenting; Vertebral augmentation; Vertebral fracture; | |
Others : 1130072 DOI : 10.1186/1471-2474-14-233 |
|
received in 2012-12-10, accepted in 2013-08-06, 发布年份 2013 | |
【 摘 要 】
Background
Up to one third of BKP treated cases shows no appreciable height restoration due to loss of both restored height and kyphotic realignment after balloon deflation. This shortcoming has called for an improved method that maintains the height and realignment reached by the fully inflated balloon until stabilization of the vertebral body by PMMA-based cementation. Restoration of the physiological vertebral body height for pain relief and for preventing further fractures of adjacent and distant vertebral bodies must be the main aim for such a method. A new vertebral body stenting system (VBS) stabilizes the vertebral body after balloon deflation until cementation. The radiographic and safety results of the first 100 cases where VBS was applied are presented.
Methods
During the planning phase of an ongoing international multicenter RCT, radiographic, procedural and followup details were retrospectively transcribed from charts and xrays for developing and testing the case report forms. Radiographs were centrally assessed at the institution of the first/senior author.
Results
100 patients (62 with osteoporosis) with a total of 103 fractured vertebral bodies were treated with the VBS system. 49 were females with a mean age of 73.2 years; males were 66.7 years old. The mean preoperative anterior-middle-posterior heights were 20.3-17.6-28.0 mm, respectively. The mean local kyphotic angle was 13.1°. The mean preoperative Beck Index (anterior edge height/posterior edge height) was 0.73, the mean alternative Beck Index (middle height/posterior edge height) was 0.63. The mean postoperative heights were restored to 24.5-24.6-30.4 mm, respectively. The mean local kyphotic angle was reduced to 8.9°. The mean postoperative Beck Index was 0.81, the mean alternative one was 0.82. The overall extrusion rate was 29.1%, the symptomatic one was 1%. In the osteoporosis subgroup there were 23.8% extrusions. Within the three months followup interval there were 9% of adjacent and 4% of remote new fractures, all in the osteoporotic group.
Conclusions
VBS showed its strengths especially in realignment of crush and biconcave fractures. Given that fracture mobility is present, the realignment potential is sound and increases with the severity of preoperative vertebral body deformation.
【 授权许可】
2013 Diel et al.; licensee BioMed Central Ltd.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20150226161436896.pdf | 1363KB | download | |
Figure 5. | 59KB | Image | download |
Figure 4. | 44KB | Image | download |
Figure 3. | 64KB | Image | download |
Figure 2. | 126KB | Image | download |
Figure 1. | 10KB | Image | download |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
Figure 4.
Figure 5.
【 参考文献 】
- [1]Muller CW, Gosling T, Mameghani A, Stier R, Klein M, Hufner T, Krettek C: [Vertebral fractures due to osteoporosis. kyphoplasty and vertebroplasty vs conservative treatment]. Orthopade 2010, 39:417-424.
- [2]Hulme PA, Krebs J, Ferguson SJ, Berlemann U: Vertebroplasty and kyphoplasty: a systematic review of 69 clinical studies. Spine (Phila Pa 1976) 2006, 31:1983-2001.
- [3]Papanastassiou ID, Phillips FM, Van Meirhaeghe J, Berenson JR, Andersson GB, Chung G, Small BJ, Aghayev K: Vrionis. Comparing effects of kyphoplasty, vertebroplasty, and non-surgical management in a systematic review of randomized and non-randomized controlled studies. Eur Spine J: FD; 2012.
- [4]Alvarez L, Alcaraz M, Perez-Higueras A, Granizo JJ, De Miguel I, Rossi RE, Quinones D: Percutaneous vertebroplasty: functional improvement in patients with osteoporotic compression fractures. Spine (Phila Pa 1976) 2006, 31:1113-1118.
- [5]Han S, Wan S, Ning L, Tong Y, Zhang J, Fan S: Percutaneous vertebroplasty versus balloon kyphoplasty for treatment of osteoporotic vertebral compression fracture: a meta-analysis of randomised and non-randomised controlled trials. Int Orthop 2011, 35:1349-1358.
- [6]Ledlie JT, Renfro MB: Kyphoplasty treatment of vertebral fractures: 2-year outcomes show sustained benefits. Spine (Phila Pa 1976) 2006, 31:57-64.
- [7]Rotter R, Martin H, Fuerderer S, Gabl M, Roeder C, Heini P, Mittlmeier T: Vertebral body stenting: a new method for vertebral augmentation versus kyphoplasty. Eur Spine J 2010, 19:916-923.
- [8]Roder C, El-Kerdi A, Eggli S, Aebi M: A centralized total joint replacement registry using web-based technologies. J Bone Joint Surg Am 2004, 86-A:2077-2079. discussion 2079–2080
- [9]Magerl F, Aebi M, Gertzbein SD, Harms J, Nazarian S: A comprehensive classification of thoracic and lumbar injuries. Eur Spine J 1994, 3:184-201.
- [10]Klezl Z, Majeed H, Bommireddy R, John J: Early results after vertebral body stenting for fractures of the anterior column of the thoracolumbar spine. Injury 2011, 42:1038-1042.
- [11]Muto M, Greco B, Setola F, Vasallo P, Ambrosanio G, Guarnieri G: Vertebral Body Stenting System for the Treatment of Osteoporotic Vertebral Compression Fracture: Follow-up at 12 Months in 20 Cases. Neuroradiol J 2011, 24:9.
- [12]Thaler M, Lechner R, Nogler M, Gstottner M, Bach C: Surgical procedure and initial radiographic results of a new augmentation technique for vertebral compression fractures. Eur Spine J 2013, 22:1608-16.
- [13]Werner CM, Osterhoff G, Schlickeiser J, Jenni R, Wanner GA, Ossendorf C, Simmen HP: Vertebral body stenting versus kyphoplasty for the treatment of osteoporotic vertebral compression fractures: a randomized trial. J Bone Joint Surg Am 2013, 95:577-584.
- [14]Rohlmann A, Zander T, Bergmann G: Spinal loads after osteoporotic vertebral fractures treated by vertebroplasty or kyphoplasty. Eur Spine J 2006, 15:1255-1264.
- [15]Feltes C, Fountas KN, Machinis T, Nikolakakos LG, Dimopoulos V, Davydov R, Kassam M, Johnston KW, Robinson JS: Immediate and early postoperative pain relief after kyphoplasty without significant restoration of vertebral body height in acute osteoporotic vertebral fractures. Neurosurg Focus 2005, 18:e5.
- [16]Voggenreiter G: Balloon kyphoplasty is effective in deformity correction of osteoporotic vertebral compression fractures. Spine (Phila Pa 1976) 2005, 30:2806-2812.
- [17]Maestretti G, Cremer C, Otten P, Jakob RP: Prospective study of standalone balloon kyphoplasty with calcium phosphate cement augmentation in traumatic fractures. Eur Spine J 2007, 16:601-610.
- [18]Kruger A, Zettl R, Ziring E, Mann D, Schnabel M, Ruchholtz S: Kyphoplasty for the treatment of incomplete osteoporotic burst fractures. Eur Spine J 2010, 19:893-900.
- [19]Heyde CE, Tschoeke SK, Hellmuth M, Hostmann A, Ertel W, Oberholzer A: Trauma induces apoptosis in human thoracolumbar intervertebral discs. BMC Clin Pathol 2006, 6:5. BioMed Central Full Text
- [20]Haschtmann D, Stoyanov JV, Gedet P, Ferguson SJ: Vertebral endplate trauma induces disc cell apoptosis and promotes organ degeneration in vitro. Eur Spine J 2008, 17:289-299.
- [21]Diel P, Freiburghaus L, Roder C, Benneker LM, Popp A, Perler G, Heini PF: Safety, effectiveness and predictors for early reoperation in therapeutic and prophylactic vertebroplasty: short-term results of a prospective case series of patients with osteoporotic vertebral fractures. Eur Spine J 2012, 21(Suppl 6):792-799.
- [22]Wardlaw D, Cummings SR, Van Meirhaeghe J, Bastian L, Tillman JB, Ranstam J, Eastell R, Shabe P, Talmadge K, Boonen S: Efficacy and safety of balloon kyphoplasty compared with non-surgical care for vertebral compression fracture (FREE): a randomised controlled trial. Lancet 2009, 373:1016-1024.
- [23]Kurth A, Bayer-Helms H, Boewe C, Hartwig E, Hoehn W, Jerosch J, Joellenbeck B, Maestretti G, Vogler W, Roellinghoff M: Radiofrequency Kyphoplasty – a novel vertebral augmentation system. Osteologie 2012, 21:4.
- [24]Georgy BA: Clinical experience with high-viscosity cements for percutaneous vertebral body augmentation: occurrence, degree, and location of cement leakage compared with kyphoplasty. AJNR Am J Neuroradiol 2010, 31:504-508.