IEEE Radio and Antenna Days of the Indian Ocean 2015 | |
RF Pulse Signal Integrity Analysis for Nonlinear Ended Microstrip Line Atom-Probe Tomography | |
无线电电子学 | |
Zhao, L.^1 ; Delamare, A.^1 ; Normand, A.^1 ; Delaroche, F.^1 ; Latry, O.^1 ; Vurpillot, F.^1 ; Ravelo, B.^2 | |
UMR CNRS 6634 - Normandy University GPM, Université and INSA de Rouen, Saint-Etienne-du-Rouvray, France^1 | |
IRSEEM EA 4353, Graduate School of Engineering, ESIGELEC, Saint-Etienne-du-Rouvray, France^2 | |
关键词: Atom probe tomography; Cryogenic chamber; Interconnect lines; Optical-controlled; Rectangular shapes; S-Parameter measurements; Signal Integrity; System frequency; | |
Others : https://iopscience.iop.org/article/10.1088/1757-899X/120/1/012006/pdf DOI : 10.1088/1757-899X/120/1/012006 |
|
来源: IOP | |
【 摘 要 】
A signal integrity (SI) analysis of high voltage rectangular short pulses for the atom- probe system is explored in this paper. The operated RF transient pulse is considered for exciting on material sample inside an ultra-high vacuum (UHV) cryogenic chamber. The ns- duration pulse signal is injected into the cryogenic analysis chamber through the transmitting system mainly constituted by a microstrip interconnect line ended by optical controlled nonlinear load. The whole system frequency characterization is performed based on the S- parameter measurements. As expected, a challenging ultra-short rectangular shape pulse is exhibited by the pulser. Promising experimental results with the improvement of ion mass spectrum is demonstrated with the designed RF pulser.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
RF Pulse Signal Integrity Analysis for Nonlinear Ended Microstrip Line Atom-Probe Tomography | 800KB | download |