On the response of electronic personal dosimeters in constant potential and pulsed x- ray beams
Guimarães, M.C.^1 ; Silva, C.R.E.^1 ; Oliveira, P.M.C.^2 ; Da Silva, T.A.^1
Centro de Desenvolvimento da Tecnologia Nuclear, CDTN, Av. Presidente Antonio Carlos 6627, Pampulha, Belo Horizonte
MG
31270-901, Brazil^1
Faculdade de Medicina, Universidade Federal de Minas Gerais, UFMG, Av. Prof. Alfredo Balena, 190, Funcionários, Belo Horizonte
MG
30130-100, Brazil^2
关键词: Constant Potential;    Electronic personal dosimeters;    International standards;    Irradiation conditions;    Pulsed radiation;    Pulsed X-rays;    Solid state detectors;    X-ray quality;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/733/1/012094/pdf
DOI  :  10.1088/1742-6596/733/1/012094
来源: IOP
PDF
【 摘 要 】

Electronic personal dosimeters (EPDs) based on solid state detectors have widely been used but some deficiencies in their response in pulsed radiation beams have been reported. Nowadays, there is not an international standard for pulsed x-ray beams for calibration or type testing of dosimeters. Irradiation conditions for testing the response of EPDs in both the constant potential and pulsed x-ray beams were established in CDTN. Three different types of EPDs were tested in different conditions in similar ISO and IEC x-ray qualities. Results stressed the need of performing additional checks before using EPDs in constant potential or pulsed x-rays.